期刊文献+

支架植入颅内蜿蜒型动脉瘤的血流动力学仿真 被引量:8

Simulation of hemodynamics in stent Intervention of basilar sinuous aneurysms
下载PDF
导出
摘要 目的构建蜿蜒型动脉瘤和弯曲支架三维有限元模型,研究支架植入动脉瘤后的血流动力学的变化。方法通过CAD软件构建出几何实体模型,借助有限元软件利用计算流体力学方法,分别对无支架和有支架的蜿蜒型脑动脉瘤定常流动血流动力学进行数值模拟,分析在动脉瘤中植入内支架前后的瘤腔内流动情形、压力和壁面切应力分布的变化情况。结果有支架动脉瘤模型上游瘤腔内的血流速度被大大削弱,圆顶突起处局部高压力明显减弱,在下游瘤腔沿壁面的压力也得到明显降低并且分布也均衡了很多,末端唇缘处局部高切应力消失了,出现的是较小且均衡的切应力。结论有支架模型瘤腔内的流速明显减小,均衡的压力分布与瘤腔内减弱的流动速度是相互统一的,利于瘤腔内血栓的形成。 Objective Study the changes of hemodynamics after stent intervention to sinuous aneurysms. Method Geometrically idealized model of three-dimensional sinuous aneurysms and curved stent was constructed using CAD software. Numerical simulations of stented and nonstented models were performed respectively in steady flow according to CFD method using finite element software. Hemodynamics data in the two models were compared in terms of flow field, pressure and the distribution of Wall Shear Stress (WSS). Result Velocity of blood flow in stented model is greatly weaken than the nonstented, and the local high pressure is markedly decreased at the proximal aneurismal dome, and the pressure along the wall of the distal aneurysm is also decreased and more uniform. The local elevated WSS atthe proximal lip of the distal an- eurysm disappeared after stent intervention, instead small and uniform WSSs distribute along a majority of the aneurismal wall. Conclusions Velocity of flow is obviously depressed after stent intervention, and the counterpoised pressure distribution of the stented model is more uniform with the weakened flow velocity in the aneurysm cavity, which are beneficial to the thrombus formation in the aneudsmal cavity.
出处 《医用生物力学》 EI CAS CSCD 2009年第1期64-69,共6页 Journal of Medical Biomechanics
基金 国家自然科学基金资助项目(10772010) 北京市自然科学基金资助项目(3062003 3092004)
关键词 血管内支架 蜿蜒型动脉瘤 计算流体力学 血流动力学 Endovascular stent Sinuous aneurysms Computational fluid dynamics Hemodynamics
  • 相关文献

参考文献20

  • 1Marks MP, Dake MD, Steinberg GK, et al. Stent placement for arterial and venous cerebrovascular disease preliminary experience [J].Neuroradiology,1994,191:441-446.
  • 2Higashida ,Randall ,SmithW,et al. Intravascular stent and endovascular coil placement for a ruptured fusiform aneurysm of the basilar artery [J]. Neurosurg ,1997,87:944- 949.
  • 3Horowitz B,Levy El, Koebbe J, et al. Transluminal stentassisted coil embolization of a vertebral confluence aneurysm technique report [J]. SurgNeurol,2001,552.91-296.
  • 4Ememann UU, Gronewoller E, Duffner FB, et al. Influence of geometric and hemodynamics parameters on aneurysm visualization during three-dimensional rotational angiography [ J ]. Am J Neuroradiol2003, 24:597-603.
  • 5Minsuok KM, Dale BT, Markus Tremmel, et al. Comparison of two stent in modifying cerebral aneurysm hemodynamics [J]. Annals of Biomedical Engineering,2008,36: 726-741.
  • 6Raghavan M, Harbaugh R. Quantified aneurysm shape and rupture risk[J]. Journal of Neurosurgery ,2005, 102: 355-362.
  • 7乔爱科,刘有军,张松.S形动脉中的血流动力学研究[J].医用生物力学,2006,21(1):54-61. 被引量:11
  • 8蔡彦,许世雄,景在平,梅志军,廖明芳,陆清声,袁良喜.腹主动脉瘤几何形态对血液动力学影响的三维数值分析[J].医用生物力学,2008,23(2):140-146. 被引量:15
  • 9Alvaro V, Sergio B,Juan S, et al. Comparison of haemodynamics in cerebral aneurysms of different sizes located in the ophthalmic artery [J]. Numer Meth Fluids, 2007,53: 793-809.
  • 10乔爱科,刘有军,张松.支架治疗主动脉弓内侧动脉瘤的仿真研究[J].生物医学工程学杂志,2007,24(4):852-856. 被引量:14

二级参考文献63

  • 1齐巍,赵继宗,王硕,王嵘.免疫炎性反应与颅内囊性动脉瘤基质重塑的组织病理学研究[J].首都医科大学学报,2004,25(4):461-463. 被引量:10
  • 2Deplano V,Bertolotti C,Boiron O.Numerical simulations of unsteady flows in a stenosed coronary bypass graft [J].Med Biol Eng Comput,2001,39:488-499.
  • 3Zhao SZ,Xu XY,Hughes AD.Blood flow and vessel mechanics in a physiologically realistic model of a human carotid arterial bifurcation[J].J Biomech,2000,33:975-984.
  • 4Moore JA,Steinman DA,Prakash S.A numerical study of blood flow patterns in anatomically realistic and simplified end-to-side anastomoses[J].J Biomech Eng,1999,121:265-272.
  • 5Moore JA,Steinman,Holdsworth DW,et al.Accuracy of computational hemodynamics in complex arterial geometries reconstructed from magnetic resonance imaging [J].Ann Biomed Eng,1999,27:32-41.
  • 6Bathe M,Kamm RD.A fluid-structure interaction finite element analysis of pulsatile blood flow through a compliant stenotic artery[J].J Biomech Eng,1999,121:361-369.
  • 7Tada S,Oshima S,Yamane R.Classification of pulsating flow patterns in curved pipes[J].J Biomech Eng,1996,118:311-317.
  • 8Stroud JS,Berger SA,Saloner D.Influence of stenosis morphology on flow through severely stenotic vessels:implication for plaque rupture[J].J Biomech,2000,33:443-455.
  • 9Naruse T,Tanishita K.Large curvature effect on pulsatile entrance flow in a curved tube:model experiment simulating blood flow in an aortic arch[J].J Biomech Eng,1996,118:180-186.
  • 10Shipkowitz T,Rokgers VG.J,Lee F J,et al.Numerical study on the secondary flow in the human aorta on local shear stresses in abdominal aortic branches[J].J Biomech,2000,33:717-728.

共引文献57

同被引文献141

引证文献8

二级引证文献68

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部