期刊文献+

影响BIS监测机械通气患者镇静深度的相关因素研究 被引量:7

The Research of Related Factors in Affecting the Bispectral Index Monitoring for the Depth of Sedation of Ventilated Patients
原文传递
导出
摘要 目的:研究影响机械通气患者BIS值的相关因素。方法:42例机械通气患者,BIS模块连续监测镇静深度48小时,记录患者镇静前后、每隔16小时的呼吸循环指标,BIS值、SAS评分,建立混合线形统计模型,评价BIS监测的影响因素。结果:患者的血压、心率、呼吸频率、PaO2、SPO2、FiO2等均对BIS值无影响,仅有PaCO2和SAS评分与BIS相关(P<0.05)。结论:BIS模块监测机械通气患者镇静深度,BIS值的变化与患者PaCO2及SAS评分有关。 Objective:To investigate the related factors affecting ventilated patients' value of bispectral index. Methods:Totally 42 ventilated patients were assigned to receive the sedation for 48 hours based on BIS monitor,the parameters of respiration, circulation, BIS value and SAS scale at before and after sedation. Every 16 hours, established the Mixed Linear Statistical Model and assessed the related factors affecting the value of bispectral index. Results: Statistically significant difference was not found in the relationship between blood pressure, heart rate, respiratory rate, PaO2, SPO2 ,FiO2 and BIS value(P〉0.05)while PaCO2 and SAS scale were related with BIS value in Statistically significant difference(P〈0.05). Conclusion: BIS model monitors the depth of sedation of ventilated patients. The variation of BIS value was related to the PaCO2 and SAS scale of ventilated patients.
出处 《华西医学》 CAS 2009年第2期362-363,共2页 West China Medical Journal
基金 成都市科学技术局2007年成都市攻关计划资助项目(成科计[2007]1号)
关键词 脑电双频指数 PACO2 SAS评分 相关因素 bispectral index PaCO2 SAS scale related factors
  • 相关文献

参考文献6

  • 1Brocas, E, Dupont, H, Paugam Burtz C. Bispectral index variations during tracheal suction in mechanically ventilated critically ill patients : effect of an alfentanil bolus[J]. Intensive Care Med, 2002,28 (2):211-213.
  • 2Ball J. How useful is the bispectral index in the management of ICU patients? [J]Minerva Anestesiol,2002,68(4) :248-251.
  • 3Simmons L E,Riker R R,Prato B S,etal. Assessing sedation during intensive care unit mechanical ventilation with the Bispectral Index and the Sedation-Agitation Scale[J]. Crit Care Med, 1999,27 (8): 1499-1504.
  • 4Mondello E, Siliotti R, Noto G, et al. Bispectral Index in ICU : correlation with Ramsay Score on assessment of sedation level[J]. J Clin Monit Comput,2002,17(5) : 271-277.
  • 5Riker R R,Fraser G L,Simmons L E,etal. Validating the Sedation- Agitation Scale with the Bispectral Index and Visual Analog Scale in adult ICU patients after cardiac surgery[J]. Intensive Care Med, 2001,27(5) :853-858.
  • 6Medicis JJ, Hu A, Lopez C, et al. Comparsion of standard sedation assessment to bispectral analysis in patients who are on neuromuscular blockers: A pilot study[C]. ASHP Midyear Clinical Meeting ( ASHP Midyear Clinical Meeting), 2002,37 (DEC) : P-347 R.