FlexRate法在AEROSET 2000全自动生化分析仪上检测酶活性的研究

Application of FlexRate method for detecting enzyme activities in Abott AEROSET 2000 automatic analyzer
下载PDF
导出
摘要 目的研究FlexRate法在雅培AEROSET2000全自动生化分析仪上检测超高酶活性样本的应用价值。方法应用FlexRate法测定丙氨酸氨基转移酶(ALT)、门冬氨酸氨基转移酶(AST)、碱性磷酸酶(ALP)、γ-谷氨酰转移酶(GGT)、肌酸激酶(CK)和淀粉酶(AMY),对FlexRate法的精密度和线性范围进行测定,并将该法测定与稀释模式测定的结果进行对比。结果FlexRate法测定ALT、AST、ALP、GGT、CK、AMY的检测上限分别提高到3 356、3 194、4 368、2 955、5 437、4 878 U/L,该法测定6种酶的批内变异系数(CV)均小于常规法。FlexRate法测定与稀释模式测定的结果比较,相关性良好。结论应用FlexRate法测定6种酶活性,重复性良好,提高了检测上限,FlexRate法测定与稀释模式测定的结果有良好的相关性。FlexRate法测定可大大减少稀释重做的次数,提高工作效率。 Objective To investigate the application value of FlexRate method for detecting enzyme activities in the Abbott AEROSET 2000 automatic analyzer. Methods The precision and linearity of FlexRate method used to detect alanine aminotransferase ( ALT), aspartate aminotransferase (AST) , alkaline phosphatase (ALP),γ-glutamyhransferase (GGT), creatine kinase(CK), and amylase (AMY) were determined. The FlexRate results were compared with the results obtained by re - analysis with dilution model. Results The upper limits of FlexRate method for detecting ALT, AST, ALP, GGT, CK,and AMY were extended to 3 356 U/L,3 194 U/L,4 368 U/L,2 955 U/L,5 437 U/L,and 4 878 U/L, respectively. The within - run coefficients of variation(CV) in FlexRate method were lower than those in routine method. The results obtained by FlexRate and re - analysis with dilution model were in good agreement. Conclusions The FlexRate method with good pre- cision extends the linearity of six enzyme assays. The correlation of FlexRate results with values obtained by dilution model are in good agreement. The re -analysis frequency with dilution model is reduced and laboratory efficiency is enhanced by using FlexRate.
出处 《武警医学》 CAS 2009年第3期206-209,共4页 Medical Journal of the Chinese People's Armed Police Force
关键词 活性 自动化分析 Enzyme activity Automatic analysis
  • 相关文献

参考文献8

  • 1Yoosoo C, Seungho R , Eunju Set al. Higher concentrations of alanine aminotransferase within the reference interval predict nonalcoholic fatty liver disease [ J ]. Clinical Chemistry,2007, 53:686
  • 2Butch AW. Dilution protocols for detection of hook effects/prozone phenomenon[ J]. Clin Chem, 2000, 46 (10) :1719
  • 3孙明洪,许志良,陈益民.雅培ARCHITECT ci8200全自动生化免疫分析仪FlexRate法检测酶活性评价[J].检验医学,2006,21(5):461-464. 被引量:2
  • 4Schwarz M J, Domke I, Helander A et al. Multicentre evaluation of a new assay for determination of carbohydrate - deficient transferrin [ J]. Alcohol, 2003,38 (3) :270
  • 5Sturm E, Albrecht - Groos R, Seyfarth M. Evaluation of new oxidation methods for the measurement of bilirubin on the aeroset clinical chemistry analyzer and comparison with methods on the Hitachi 717 [ J ]. Clin Lab, 2002,48 (7 -8) :401
  • 6National Committee for Clinical Laboratory Standards. Evaluation of the linearity of quantitative analytical methods ; proposed guideline ( EP6 - P) [ S ]. Villanova, PA : NCCLS, 1986
  • 7Pauli D, Seyfarth M, Dibbelt L. The Abbott Architect c8000: analytical performance and productivity characteristics of a new analyzer applied to general chemistry testing[ J ]. Clin Lab, 2005,51 ( 1 - 2) :31
  • 8Duly EB, Barnes G, Grimason S, Analytical Performance of Specific - Protein Assays on the Abbott Aeroset System [J]. Clin Chem, 2001,47:1709

二级参考文献3

  • 1National Committee for Clinical Laboratory Standards.Evaluation of the linearity of quantitative analytical methods; proposed guideline (EP6-P)[ S ].Villanova,PA:NCCLS,1986.
  • 2National Committee for Clinical Laboratory Standards.Method comparison and bias estimation using patient samples; approved guideline (EP9-A)[ S ].Wayne,PA:NCCLS,1995.
  • 3United States Patent 5025389.Method of analyzing reaction rate in Chemical analysis[ S].1991.

共引文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部