期刊文献+

钼盐涂覆-无火焰原子吸收法测定肿瘤患者血清中微量铝的研究 被引量:2

DETERMINATION OF TRACE Al IN SERUM OF CANCER PATIENTS BY GRAPHITE FURNACE ATOMIC ABSORPTION SPECTROMETRY USING K2Cr2O7 AND SURFACTANT AS MATRIX MODIFIER WITH MOLYBDENATE COATED TUBES
下载PDF
导出
摘要 用钼盐涂覆石墨管并以K2Cr2O7和TritonX-100为基体改进剂,GFAAS法测定肿瘤患者血清中微量铝,大幅度提高测定灵敏度和准确度。避免了沾污。特征浓度21.15pg/0.0044AS。加标平均回收率104.92%。相对标准偏差小于8.17%。 Different kinds of graphite tubes and of matrix modifiers have been tested for optimization of operation condition for determination of Al in serum by GFAAS.The effects of different interference ions have been eliminated by adding K2Cr2O7 and Triton X100 as matrix modifier.The method is suitable for determination of serum of cancer patients.The characteristic mass is 2115pg/00044A·S.The recovery rate is 10492.The relative standard deviation is less than 817%.The result obtained is satisfactory.
作者 卜海富 孙昕
出处 《光谱学与光谱分析》 SCIE EI CAS CSCD 北大核心 1998年第1期81-84,共4页 Spectroscopy and Spectral Analysis
关键词 血清 钼涂覆管 GFAAS 肿瘤患者 Serum, Molybdenate coated tubes, GFAAS, Al