摘要
Synchrotron X-ray fluorescence analyses of crystal and amorphous clinopyroxene were compared. The results showed that using the synchrotron X-ray as a source of energy, the diffraction X-ray of crystal materials will seriously affect the X-ray fluorescence analysis. In order to avoid the influence of diffraction, the best way to solve it is to use the monochromatic source, or to have a collimating slit between sample and Si(Li) detector.
Synchrotron X-ray fluorescence analyses of crystal and amorphous clinopyroxene were compared. The results showed that using the synchrotron X-ray as a source of energy, the diffraction X-ray of crystal materials will seriously affect the X-ray fluorescence analysis. In order to avoid the influence of diffraction, the best way to solve it is to use the monochromatic source, or to have a collimating slit between sample and Si (Li) detector.