期刊文献+

A double-parameter theoretical model of the conduction calorimeter

A double-parameter theoretical model of the conduction calorimeter
原文传递
导出
摘要 <正> A double-parameter theoretical model for the conduction calorimeter, in which the delay of sensing has been considered, is suggested, and a simple method for the determination of the parameters of the model is proposed. The experimental results show that on the basis of this model, the phenomenon of thermogram maximum in thermogram reconstruction can be eliminated, and it can be applied to the thermokinetic investigations of faster thermal change processes. A double-parameter theoretical model for the conduction calorimeter, in which the delay of sensing has been considered, is suggested, and a simple method for the determination of the parameters of the model is proposed. The experimental results show that on the basis of this model, the phenomenon of thermogram maximum in thermogram reconstruction can be eliminated, and it can be applied to the thermokinetic investigations of faster thermal change processes.
出处 《Science China Chemistry》 SCIE EI CAS 1995年第6期658-667,共10页 中国科学(化学英文版)
基金 Project supported by the National Natural Science Foundation of China.
关键词 conduction CALORIMETER THEORETICAL model thermogram MAXIMUM CALORIMETRY thermokinetics. conduction calorimeter, theoretical model, thermogram maximum, calorimetry, thermokinetics.