期刊文献+

血清ALT紫外监测法试剂盒的研制和应用

DEVELOPMENT AND APPLICATION OF A ALT(GPT) KIT BASED ON UV-MONITORING METHOD
原文传递
导出
摘要 作者采用国产原料研制成功适用于在临床生化自动分析仪或紫外分光光度计上测定血清ALT(GPT)活性的ALT紫外监测法试剂盒.在Impact400型生化仪上所作的试剂盒性能质量评价实验证明,酶促反应的延迟时间短(<20s),批内变异系数1.1~6.5%,血清ALT活性线性可达1000u/L(37℃),与进口同类试剂盒对比测定结果一致,呈非常显著相关.自制ALT试剂盒4~8℃存放可稳定6个月.81名健康成人血清ALT活性为26.91±8.91u/L(37℃). A kinetic ALT(GPT) kit,based on UV-monitoring method,was developed for the determination of serum alanine aminotransferase(ALT/GPT) with automated analyzer.Methodological studies on Impact 400 analyzer showed that the coefficient of variation within run was 1.1-6.5%;the linearity between serum ALT activity and amount of serum used in the reaction system was up to 1000μ/L (37℃);the lag phase in the reaction system was less than 20 sec;a good correlation between the present kit and those of Boehringer Mannheim (W.Germany) and Gilford (USA) was obtained.The present kit kept stable for 6 months when stored at 4-8℃.Serum ALT activity (?±SD) of 81 healthy adults was 26.91 ±8.91μ/L (37℃).
作者 沈伽弟
出处 《军事医学》 CAS CSCD 北大核心 1989年第4期277-281,共5页 Military Medical Sciences
基金 军事医学科学院科学基金(科基字022号)
关键词 丙氨酸氨基移换酶(谷丙转氨酶) 紫外监测法 临床生化自动分析仪 试剂盒 alanine aminotransferase reagent kit UV-monitoring method automated clinical chemistry analyzer
  • 相关文献

参考文献3

  • 1呼延铃,戴起云,解瑜,王金良.国产试剂盒在CORONA生化分析仪上应用的评价[J]天津医药,1988(01).
  • 2王毓三.酶动力学测定原理与方法[J]临床检验杂志,1985(04).
  • 3赵善政,杨伟宗,贾维龙,沈斐纳,邱鹤钧,吴伯惠.血清谷丙转氨酶(SGPT)活力的酶偶联分光光度测定法[J]临床检验杂志,1985(01).