摘要
介绍了REVOLUTION^(TM)XR/d DR系统故障维修,故障现象间断出现,由于设备系统内部通讯与外部通讯混用交换机,导致系统不能正常启动。通过改变连接,系统内部的控制通讯不受外部干扰,设备工作正常。
In this paper, the case of maintaining the malfunction of REVOLUTIONTM XR/d DR system was described. Breakdowns happened now and then, and the system could not start properly as a result of the mixed use of interior communication and exterior communication. To make it back to normal, connectors have been changed and there was no exterior interference now.
出处
《中国医学装备》
2009年第4期55-56,共2页
China Medical Equipment
关键词
系统
故障维修
平板探测器
System
Troubleshooting
Flat panel detector