FW型轴流泵的体外溶血与动物实验研究 被引量:14

Experimental Study of Hemolysis in Vitro and Animal Testing for Fu Wai Axial Blood Pump
下载PDF
导出
摘要 目的为给心力衰竭患者提供有效的心室机械辅助治疗,研制出可以部分或完全代替心室功能的FW(FuWai)型轴流式心室辅助血泵,评价其流体力学性能和血液相容性能。方法通过体外测试台和5只小尾寒羊的动物实验,对FW型轴流式心室辅助血泵的流体力学性能和血液相容性能进行了测试。结果轴流泵在转速为8400r/min、后负荷为100mmHg的情况下,流量可达5L/min;体外溶血试验检测结果表明轴流泵的标准溶血指数(NIH)为0.17±0.06mg/L。动物实验结果表明,血浆游离血红蛋白含量在30mg/dl左右。结论FW型轴流式心室辅助血泵的血液相容性能良好,体外和体内试验结果达到了国际标准,有望在完善动物实验的基础上进入临床试验阶段。 Objective To provide a ventricular assist device for patients with heart failure, Fu Wai (FW) axial blood pump was developed for partly or totally to assist the left ventricular function. Vitro hemolysis and animals tests were also employed to test the hydromechanics and hemocompatibility of the FW left ventricular assist devices developed in Fu Wai hospital. Methods Using vitro test loop, FW axial blood pump has been used to evaluate the performance of hemolysis, the pump has also been tested for hemolysis characteristic through five sheep experiments. Results At 8 400 r/min, the pump generates 5 L/min flow against 100 mm Hg, the normalized index of hemolysis (NIH) was 0.17±0.06 mg/L. The plasma free hemoglobin of in vivo tests was around 30 mg/dl. Conclusion The results obtained in vitro and in vivo testing indicate an acceptable design for the blood pump, further in vivo tests will be performed before clinical use.
出处 《中国胸心血管外科临床杂志》 CAS 2009年第2期114-117,共4页 Chinese Journal of Clinical Thoracic and Cardiovascular Surgery
基金 国家十一五课题基金资助项目(2003BAI01A00) 863课题资助项目(2006AA02Z4D3)~~
关键词 轴流泵 溶血 体外试验 动物试验 标准溶血指数 Axial blood pump Hemolysis In vivo testing Animal testing Normalized index of hemolysis
  • 相关文献

参考文献9

  • 1Chinn R, Dembitsky W, Eaton L, et al. Multicenter experience: prevention and management of left ventricular assist device infections. ASAIOJ, 2005, 51(4):461-470.
  • 2Haddad H, Elabbassi W, Moustafa S, et al. Left ventricular assist devices as bridge to heart transplantation in congestive heart failure with pulmonary hypertension. ASAIO J, 2005, 51 (4) :456-460.
  • 3Davies RR, Russo MJ, Hong KN, et al. The use of mechanical circulatory support as a bridge to transplantation in pediatric patients:an analysis of the United Network for Organ Sharing database. J ThoracCardiovasc Surg, 2008, 135(2) :421-427.
  • 4Anyanwu AC, Fischer GW, Plotkina I, et al. Off-pump implant of the Jarvik 2000 ventricular assist device through median sternotomy. Ann ThoracSurg, 2007, 84(4):1405-1407.
  • 5Zhang Y, Xue S, Gui XM, etal. A novel integrated rotor of axial blood flow pump designed with computational fluid dynamics. Artif Organs, 2007, 31(7) :580-585.
  • 6Zhang Y, Zhan Z, Gui XM, et al. Design optimization of an axial blood pump with computational fluid dynamics. ASAIO J, 2008, 54(2):150-155.
  • 7张岩,薛嵩,桂幸民,胡盛寿,孙寒松,朱晓东.运用三维数值模拟对人工心脏轴流血泵的设计和改进[J].中国生物医学工程学报,2007,26(1):35-41. 被引量:20
  • 8张岩,孙寒松,周建业,张浩,郑哲,李国荣,朱晓东,桂幸民,胡盛寿.一种新型轴流式心脏辅助血泵的研制和初步的动物实验[J].中国生物医学工程学报,2008,27(1):97-101. 被引量:10
  • 9Haft J, Armstrong W, Dyke DB, et al. Hemodynamic and exercise performance with pulsatile and continuous-flow left ventricular assist devices. Circulation, 2007, 116 ( 11 Suppl) :I8-I15.

二级参考文献25

  • 1Hetzer R, Muller J, Weng Y, et al. Cardiac recovery in dilated cardiomyopathy by unloading with a left ventricular assist device[J].Ann Thorac Surg, 1999, 68:742 - 749.
  • 2Frazier OH, Delgado RM. Mechanical circulatory support for advanced heart failure. Where does it stand in 2003? Circulation[J], 2003,108:3064 - 3068.
  • 3Rose EA, C, elijns AC, Moskowltz AJ, et al. Long-term use of a left ventricular assist device for end-stage heart failure [ J ]. N Engl J Med, 2001,345:1435 - 1443.
  • 4Song XW, Throckmorton A L., Untaroiu A, et al. Axial flow blood pumps[J]. ASAIO J, 2003, 49:355 - 364.
  • 5Untaroiu A, Throckmorton AL, Patel SM, et al. Numerical and experimental analysis of an axial flow left ventricular assist device:the influence of the diffuser on overall pump performance [ J ].Artificial Organs, 2005, 29(7): 581 -591.
  • 6Chan WK, Wong YW, Ong W, et al. Numerical investigation of the effects of the clearance gap between the inducer and impeller of an axial blood pump [ J ]. Artificial Organs, 2005, 29 ( 3 ) : 250 -258.
  • 7Yamane T, Ikeda T, Orita T, et al. Fluid dynamics of turbo pumps for artificial hearts [ J]. Materials Science and Engineering C4(1996) :99- 106.
  • 8Mitoh A, Yano T, Sekine K, et al. Computational Fluid Dynamics Analysis of an Intra-Cardiac Axial Flow Pump. Artificial Organs,2003,27( 1 ) :34 - 40.
  • 9Datum G, Mizuguchi K, Bozeman R,et al. In vitro performance of the Baylor/NASA axial flow pump. Artif Organs. 1993,17(7):609- 13.
  • 10Mizuguchi K, Datum GA, Bozeman RJ, et al Development of the Baylor/NASA axial flow ventricular assist device: in vitro performance and systematic hemolysis test results[J]. Artif Organs,1994,18(1) :32 - 43.

共引文献26

同被引文献177

引证文献14

二级引证文献53

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部