Siemens primus E医用直线加速器的剂量校准 被引量:6

Dose Determination and Calibration of Siemens primus E LINAC
下载PDF
导出
摘要 介绍了用IAEA法对Siemens primus E加速器输出剂量的计算、测量和校准程序,寻求准确、快捷、有效的测量方法。按IAEA规程确定各测量参数值,将吸收剂量计算公式化简成仪表读数乘以照射量校准因子Nx和温度、气压修正因子Ktp以及总校正因子Cf的形式。通过对primus加速器的输出剂量测量并校准,使其精度符合质量保证和质量控制要求。此测量与校准方法,可以用于使用数字化医用加速器实施精确放疗对输出剂量的精度要求和临床治疗。 The procedures for the calculation and measurement and calibration of output dose of the Siemens primus E accelerator with IAEA regulation are introduced to probe an accurate, rapid and effective method .The parameter value of measurement is determined according to IAEA regulations and absorbed dose calculation formal is briefed to the form that the reading of instrument multiplies exposure calibration coefficient N, and Air temperature, pressure and humidity effects K, and cumulative correction factor Cf the precision of accelerator conform to the standard of QA&QC by the measurement and calibration of the Hospital Primus E accelerator in output dose. The methods of measurement and calibration can be used for the determination of output dose in precise radiotherapy of digital medical linear accelerator and clinical treatment.
出处 《医疗卫生装备》 CAS 2009年第4期115-117,共3页 Chinese Medical Equipment Journal
关键词 加速器 剂量校准 Cf常数 accelerator , dose calibration Cf constant
  • 相关文献

参考文献3

二级参考文献14

  • 1陈泽龙.医用回旋加速器原理[J].医疗卫生装备,2005,26(2):55-56. 被引量:10
  • 2王京陵.医用加速器调制器故障判断要点[J].医疗卫生装备,2005,26(3):67-68. 被引量:3
  • 3杨树欣,赵庆军,刘成红.加速器X-线半影的影响因素分析[J].医疗卫生装备,2005,26(6):38-38. 被引量:5
  • 4陈泽龙.医用回旋加速器及正电子药物合成系统的安全与防护[J].医疗卫生装备,2005,26(6):42-44. 被引量:7
  • 5IAEA Technical Reports Series No.277. Absorbed dose determination in photon and electron beam.An International Code of Practice,Second Edtion,IAEA,Vienna:1997.5-6.
  • 6IAEA Technical Reports Series No.277. Absorbed dose determination in photon and electron beam.An International Code of Practice,Second Edtion,IAEA,Vienna:1997.39.
  • 7IAEA Technical Reports Series No.277. Absorbed dose determination in photon and electron beam.An International Code of Practice,Second Edtion,IAEA,Vienna:1997.54-55.
  • 8Saiful MH.Calibration of high-energy photon and electron beams for radiotherapy using AAPM.1983 and IAEA.1987 dosimetryprotocols. Med Phys,1993,20:293.
  • 9IAEA Technical Reports Series No.277. Absorbed dose determination in photon and electron beam.An International Code of Practice,Second Edtion,IAEA,Vienna:1997.25.
  • 10国家质量技术监督局.中华人民共和国国家计量检定规程:外照射治疗辐射源.2001

共引文献6

同被引文献43

引证文献6

二级引证文献12

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部