摘要
在H2SO4介质中,KMnO4氧化左羟丙哌嗪产生化学发光,甲醛可增敏上述化学发光,在此基础上建立了反相高效液相色谱(RP-HPLC)分离,柱后化学发光检测左羟丙哌嗪的新方法,并成功应用于血清中左羟丙哌嗪的测定。在优化的实验条件下,该法测定左羟丙哌嗪的线性范围为1.0×10^-7—1.0×10^-5g/mL,检出限(3σ)为3×10^-8g/mL。对2.0×10^-6g/mL的左羟丙哌嗪进行11次平行测定,其相对标准偏差为2.3%。
A novel method was developed for the determination of levodroproizine by high performance liquid chroma-tography (HPLC) with chemiluminescence (CL) detection. The procedure was based on the chemiluminescence reaction of levodroproizine and potassium permanganate with formaldehyde as an enhancement reagent in a sulfuric acid medium. The chromatographic separation was performed on a nucleosil RP-C18 column (i.d. : 250 mm×4.6 mm, particle size: 5μm, pore size: 100) with an isocratic mobile phase consisting of methanol and 0.1 mol/L NaH2PO4(volume ratio 45:55) at a flow rate of 1.0 mL/min. Under optimum conditions, the linear range for levodroproizine was 1.0×10^-7—1.0×10^-5 g/mL with the detection limit of 3×10^-8 g/mL(3σ). The relative standard deviation (RSD) was 2.3% (n= 11) for 2.0×10^-6g/mL levodroproizine. Levodroproizine in the serum samples were analyzed by the proposed method with satisfactory results.
出处
《分析试验室》
CAS
CSCD
北大核心
2009年第5期16-19,共4页
Chinese Journal of Analysis Laboratory
基金
国家自然科学基金(30470886)项目资助
关键词
高效液相色谱
化学发光
左羟丙哌嗪
KMNO4
甲醛
High performance liquid chromatography
Chemiluminescence
Levodropropizine
Potassium permanga-nate
Formaldehyde