期刊文献+

高灵敏度γ剂量率监测仪的设计与研制 被引量:7

Development of Miniature γ Dose Rate Monitor with High Sensitivity
下载PDF
导出
摘要 介绍了一种基于单片机设计的小型高灵敏度γ剂量率监测仪,可对γ射线剂量率进行连续监测的同时,根据具体环境情况采用有线、无线两种通讯方式将监测数据发送给主机。仪器有交流和电池两种供电方式,可根据现场情况选用。文章详细介绍了仪器的系统结构,软、硬件各部分的设计思路和实现技术。经过应用实验表明,仪器的性能指标为:测量范围0.01μSv/h^20μSv/h,γ射线的灵敏度为900cps/μSv.h-1,死时间小于200μs,稳定性误差小于±10%。 this paper introduces a miniature γ dose rate monitor with high sensitivity which design based on single chip microcomputer, it can continue monitoring γ dose rate and then choose wire or wireless communications to sent the monitoring data to host according to the actual conditions. It has two kinds of power supply system, AC power supply system and battery which can be chose by concrete circumstances. The design idea and implementation technology of hardware and software and the system structure of the monitor are detailed illustrated in this paper. The experimental results show that measurable range is 0. 1mR/h-200mR/h, the sensitivity of γ is 90eps/mR. h-1, dead time below 200us, error of stability below t 10 %.
机构地区 成都理工大学
出处 《核电子学与探测技术》 CAS CSCD 北大核心 2009年第2期312-315,共4页 Nuclear Electronics & Detection Technology
关键词 Γ剂量率 GM计数管 高灵敏度 γ dose rate, GM tube, High sensitivity
  • 相关文献

参考文献8

二级参考文献15

  • 1张松寿.GM计数管脉冲供电电压对计数率的影响[J].核电子学与探测技术,1997,17(3):211-214. 被引量:4
  • 21,Osvay M,Stenger V,Foldiak G.Report IAEA-SM-193/60 Vienna:IAEA,1975.623-632
  • 32,Rikner G,Grusell E.Phys Med Biol,1987,32(9):1109-1117
  • 43,Dixon R L,Ekstrand K E.Nucl Instr Meth,1980,175:112-114
  • 54,Dixon R L,Ekstrand K E.Int J Appl Radiat Isot,1982,33:1171-1176
  • 6GB/T 13161-91,直读式个人X和γ辐射剂量当量和剂量率监测仪.
  • 7EJ 269-84,X、γ射线外照射个人剂量监测规定.
  • 8EJ 943-95,辐射工作人员个人监测管理规定.
  • 9ZP1210,ZP1304盖革-弥勒计数管数据手册[Z].
  • 10KS0108,点阵液晶控制器数据手册[Z].