用二维电离室阵列对螺旋断层治疗的调强计划进行剂量验证 被引量:9

Two-dimensional ion chamber array in dose verification for intensity modulated planning of helical tomotherapy
原文传递
导出
摘要 目的研究采用二维电离室阵列对螺旋断层治疗(HT)的调强计划实施剂量验证的可行性,寻求建立一套临床上针对该条件下患者治疗更为有效的剂量验证方法。方法采用IBA公司I’mRT MatriXX二维电离室阵列及其相配套MULTICube等效固体水模体对10例患者HT的调强计划实施验证。分别对二维电离室阵列实行冠状及纵向位测量,并获取模体中阵列轴平面和纵断面剂量分布。通过HT系统兆伏级CT图像实现模体精确配准及校正,以确保二维电离室阵列摆位准确性。束流照射后将二维阵列剂量测量平面分布与HT计划系统模体计划中计算平面结果进行比较,定性或定量分析其绝对剂量及相对剂量验证情况。探讨其不同位置摆放来实现其测量方法的可行性。结果定性或定量分析所测量与计算绝对剂量点及相对剂量分布的结果均显示出了较好一致性,点绝对测量与计算剂量偏差保持在±3%以内。MatriXX阵列测量的相对剂量分布与治疗计划系统模体中计算平面相比较,采用了Gamma法(3mm或3%)进行2mm栅格精度分析,1≤1像素点平均通过率分别为97.76%、96.83%。结论MatriXX二维电离室阵列可较好地进行绝对剂量及相对剂量测量,能较好地实现调强计划的剂量验证。 Objective To investigate the feasibility of dose verification of intensity modulated (IM) planning of helical tomotherapy (HT) using two-dimensional ion chamber array (2DICA), and develop an efficient way to validate the dose delivered under the parameters mirroring those during the treatment. Methods A 2DICA,I'mRT MatriXX and MULTICube equivalent solid water phantom from IBA company were used to verify the dose distribution of 10 IM planning. The combined phantom was set up to measure the dose distributions on coronal and sagittal surface. The precise setup of phantom was guided by HTMVCT images. After the irradiation, the measured dose distributions on the coronal and sagittal plane were compared with those calculated by the IM planning system for verification. The results were evaluated and the feasibility of the different measuring methods was studied. Results The dose distribution measured by the MatriXX 2DICA was well consistent with that calculated by the treatment planning system. The errors between the measured dose and predicted dose in the selected points were within ± 3%. In the comparison of the pixelsegmented ionization chamber versus treatment planning system using the 3 mm/3% γ criteria, the passing ratio of pixels with γ parameter ≤1 was 97.76% and 96.83% ,respectively. Conclusions MatriXX is able to measure the absolute and relative dose distributions simultaneously, which can be used for dose verification of IM planning.
出处 《中华放射肿瘤学杂志》 CSCD 北大核心 2009年第3期233-236,共4页 Chinese Journal of Radiation Oncology
关键词 螺旋断层治疗 二维电离室阵列 等效固体水模 剂量验证 Helical tomotherapy Two-dimensional ion chamber array Equivalent solid water phantom Dose verification.
  • 相关文献

参考文献7

  • 1Woo M, Nico A. Using a matrix detector for rotational delivery QA. Med Phys ,2007 ,34 :2502.
  • 2Amerio S, Boriano A, Bourhaleb F, et al. Dosimetric characterization of a large area pixel-segmented ionization chamber. Med Phys ,2004,31:414-420.
  • 3Low DA, Dempsey JF. Evaluation of the gamma dose distribution comparison method. Med Phys ,2003,30 : 2455-2464.
  • 4Nelms BE, Simon JA. A survey on planar IMRT QA analysis. J Appl Clin Med Phys,2007 ,8 :876-890.
  • 5Koch N,Ashenafi M,Li J,et al. Evaluation of a 2D ion chamber array for Quality Assurance of Intensity Modulated Arc Therapy. Med Phys,2008,35 : 2945.
  • 6Araki F, Tajiri S, Tominaga H, et al. Application of a 2-D ion chamber array for dose verification of dynamic IMRT with a micro- MLC. Int J Radiat Oncol Biol Phys ,2006,66:703-704.
  • 7Stasi M, Giordanengo S, Cirio R, et al. D-IMRT verification with a 2D pixel ionization chamber: dosimetric and clinical results in head and neck cancer. Phys Med Biol,2005 ,50 :4681-4694.

同被引文献131

  • 1柏朋刚,李奇欣,张秀春,费召东,缪龙华.两种验证工具对鼻咽癌调强验证结果对比分析[J].中国医学物理学杂志,2011,28(6):2964-2966. 被引量:10
  • 2Thomas SD, Mackenzie M, Field GC, et al. Patient specific treatment verifications for helical tomotherapy treatment plans. Med Phys,2005 ,32 :3793-3800.
  • 3Xu SP, Deng XW, Dai XK, et al. Quality assurance of helical tomotherapy intensity mtMulated radiation therapy. IFMBE Proceeding ,2008,19:447-450.
  • 4Saminathan S, Manickan R, Chandraraj V, et al. Dosimetric study of 2D ion chamber array matrix for the modern radiotherapy treatment verification. J Appl C l in Med Phys,2010,11 : 116-127.
  • 5Herzen J, Todorovic M, Cremers F, et al. Dosimetric evaluation of a 2D pixel ionization chamber for implementation in clinical routine. Phys Med Bio1,2007,52 : 1197-1208.
  • 6Woo M, Nico A. Using a matrix detector for rotational delivery QA. Med Phys,2007,34:2502.
  • 7Low DA, Dempsey JF. Evaluation of the gamma dose distribution comparison method. Med Phys,2003 ,30 :2455-2464.
  • 8Buonamici FB, Compagnucci A, Marrazzo L, et al. An intercomparison between film dosimetry and diode matrix for IMRT quality assurance. Med Phys,2007,34 : 1372-1379.
  • 9Han Z, Ng SK, Bhagwat MS, et al. Evaluation of MatriXX for IMRT and VMAT dose verifications in peripheral dose regions. Med Phvs 2010,37 3704-3714.
  • 10Koch N, Ashenfi M, Li J, et al. Evaluation of 2D ion chamber array for quality assuance of intensity modulated arc: therapy. Med Phys ,2008,35:2945.

引证文献9

二级引证文献48

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部