摘要
目的建立土壤中总汞的原子荧光测定方法。方法对方法中酸消解、原子荧光光度计测定条件等进行优化。结果在最佳条件下,方法的检出限为(0.25×10-3)mg/kg(按称取1g样品,定容至50ml计算),回收率为93.8%~103%,相对标准偏差为<5%。结论方法具有灵敏度高,检出限低,共存元素干扰少,线性范围宽,方法快速、简便等特点。
[ Objective] To establish a determining method of Mercury in soil by Atomic Fluorescence Spectrophotometry. [ Methods ] Acid digestion of this method, determination conditions of Atomic Fluorescence Spectrometer were optimized. [ Results ] Under optimum condition, the detection limit was (0.25 × 10^-3) mg/kg (calculation by 1 g sample to 50 ml constant volume), the coefficient of recovery was 93.8% - 103%, relatively standard deviation was 〈 5 %. [ Conclusion ] This method has high sensitivity, low detection limit, less interference with the coexistence of elements, wide linear range, rapid method, simple, and so on.
出处
《职业与健康》
CAS
2009年第11期1147-1148,共2页
Occupation and Health
关键词
土壤
总汞
原子荧光法
酸消解
Soil
Total Mercury
Atomic fluorescence spectrometry
Acid digestion