摘要
将流动注射技术引入N-(1-萘基)乙烯二胺光度法测定NO-2分析体系,采用单阀双带注入镀镉锌片还原柱带隔离的阀体流路,建立了同时测定微量NO-2和NO-3的分析方法.NO-2-N的测定范围为0.01—2.0μg/ml,NO-3-N的测定范围为0.05—5.0μg/ml,分析速度为30样/h.方法应用于环境水样及蔬菜中NO-2和NO-3的同时测定,结果满意.
A flow injection spectrophotometry is proposed for simultaneous determination of nitrite and nitrate based on the N-(1-naphthyl)- ethylenediamine spectrophotometry.A technique of double zone injection and reductor column of cadmium plated zinc power zone isolation is adopted.The ranges of determination are 0.01—2.0μg/ml for NO_2^- N and 0.05—5.0 μg/ml for NO_3^-NThe sampling rate is 30 samples per hour.This method has been applied to the determination of nitrite and nitrate in environmental waters and vagetables with satisfactory results.
出处
《环境科学》
EI
CAS
CSSCI
CSCD
北大核心
1998年第3期84-86,共3页
Environmental Science
关键词
流动注射分析
环境水样
蔬菜样
光度法
水质监测
flow injection analysis,environmental water analysis,vagetable analysis,nitrite,nitrate,simultaneous spectrophotometry.