期刊文献+

表面磁性膜医用316L不锈钢支架的生物相容性 被引量:1

Biocompatibility of surface magnetic film medical 316L stainless steel stent in vivo
原文传递
导出
摘要 目的研究表面磁性膜血管内支架的生物相容性,为该支架的临床应用提供实验依据。方法通过溶血实验、动态凝血时间实验、急性全身毒性实验、皮内刺激实验、细胞毒性实验、热源实验、过敏实验及体内植入实验,综合评价表面磁性膜血管内支架的生物相容性。结果表面磁性膜血管内支架无溶血反应及凝血功能的改变,无急性全身毒性反应,无热源反应,支架材料中不存在致敏性物质;支架材料动物体内植入在初期有轻度的炎性反应,12周后炎性反应基本消失,未见炎性细胞浸润积聚现象。结论表面磁性膜血管内支架具有良好的生物相容性,其应用于临床具有可行性和安全性。 Objective To investigate the biocompatibility and biological safety of surface magnetic film 316L medical stainless steel stent in vivo inorder to provide experimental reference for clinical application. Methods Acute systemic toxicity test, intracutaneous stimulation test, hemolysis test, cytotoxicity test, pyrogen test, sensitivity test and muscle embedding test were performed to evaluate the biocompatibility and biological safety of surface magnetic film 316L medical stainless steel stent. Results There was no potential allergic substance in magnetic stent. No hemolytic reaction, acute systemic toxicity effect or pyrogen reaction was observed. Mild inflammatory reaction was seen in the surrounding tissues in the early implantation stage, which completely disappeared after 12 weeks. Macrophage aggregation was not noted. Conclusion Surface magnetic film 316L medical stainless steel stent has fine biocompatibility, which is feasible and safe for clinical application.
出处 《中华生物医学工程杂志》 CAS 2008年第6期459-462,共4页 Chinese Journal of Biomedical Engineering
基金 国家自然科学基金(30470486) 中国博士后科学基金面上资助金(20070410383) 中国博士后科学基金特别资助金(200801439)
关键词 磁性膜 生物相容性 支架 316L不锈钢 毒性反应 炎性反应 Magnetic Biocompatibility Stent 316L stainless steel Toxic reaction Inflammatory reaction
  • 相关文献

参考文献7

  • 1Movahed MR. Coronary artery bifurcation lesion classifications, interventional techniques and clinical outcome. Expert Rev Cardiovasc Ther, 2008, 6: 261-274.
  • 2Boden WE, Hoekstra J, Miller CD. ST-elevation myocardial infarction: the role of adjunctive antiplatelet therapy. Am J Emerg Med, 2008, 26: 212-220.
  • 3Ward MR, Stewart D J, Kutryk MJ. Endothelial progenitor cell therapy for the treatment of coronary disease, acute MI, and pulmonary arterial hypertension: current perspectives. Catheter Cardiovasc Interv, 2007, 70: 983-998.
  • 4胡涛,吕安林,张荣庆,李兰荪,贾国良,曹艳杰.低频电磁场对培养的人脐静脉内皮细胞一氧化氮合酶的影响(英文)[J].中国临床康复,2004,8(12):2376-2377. 被引量:1
  • 5Hu T, Li H, Zhang XJ, et al. Exposure of aortic vascular smooth muscle cells to low- frequency electromagnetic field inhibits osteopontin expression and matrix metalloproteinase activity. International Heart Journal, 2008, 49: 597-604.
  • 6杨治军,周廉,李争显,于振涛,罗丽娟,蔡玉荣.医用316L不锈钢表面沉积Fe/Pt和Pt/Fe/Pt磁性薄膜的研究[J].稀有金属材料与工程,2007,36(11):1960-1963. 被引量:3
  • 7Murray PE, Garcia-Godoy C, Gareia-Godoy F. How is the biocompatibilty of dental biomaterials evaluated? Med Oral Patol Oral Cir Bucal, 2007, 12: E258-E266.

二级参考文献17

共引文献2

同被引文献11

  • 1Byrne RA, Joner M, Kastrati A. Polymer coatings and delayed arterial healing following drug-eluting stent implantation [J] . Minerva Car- dioangiol, 2009, 57 (5) : 567-584.
  • 2Cuneo AA, Autieri MV. Expression and function of anti-inflammatory interleukins: the other side of the vascular response to injury [J] . Curt Vasc Pharmacol, 2009, 7 (3) : 267-276.
  • 3Tao H, Huan L, Xiaojun Z, et ol. Exposure of aortic vascular smooth muscle cells to low-frequency electromagnetic field inhibits osteopon- tin expression and matrix metalloproteinase activity [J] . International Heart Journal, 2008, 49 (5) : 597-604.
  • 4Montone RA, Ferrante G, Baca M, et al. Predictive value of C-reactive protein after drug-eluting stent implantation [J] . Future Cardiol, 2010, 6 (2) : 167-179.
  • 5Hailong GE, Yujie Z, Hanying M, et al. Influence of drug-eluting stent on inflammation during restenosis in a porcine coronary model [J] . Acta Cardiol, 2010, 65 (1) : 31-35.
  • 6Saleh N, Kovacs A, Tornvall P. Relevance of genetic polymorphisms in inflammatory response to percutaneous coronary intervention [ J] . Scand J Clin Lab Invest, 2009, 69 (7) : 736-740.
  • 7Busnelli M, Froio A, Bacci ML, et al. Pathogenetic role of hypercholesterolemia in a novel preclinical model of vascular injury in pigs [J] . Atherosclerosis, 2009, 207 (2) : 384-390.
  • 8Punchard MA, O'Cearbhaill ED, Mackle JN, et ol. Evaluation of human endothelial cells post stent deployment in a cardiovascular simulator in vitro [J] . Ann Biomed Eng, 2009, 37 (7) : 1322-1330.
  • 9Zurakowski A, Wojakowski W, Dzielski T, et al. Plasma levels of C- reactive protein and interleukin-10 predict late coronary in-stent restenosis 6 months after elective stenting [J] . Kardiol Pol, 2009, 67(6) : 623-630.
  • 10胡涛,周廉,介万奇,王海昌,于振涛,李争显,李寰,杨治军.恒磁场对大鼠主动脉平滑肌细胞骨桥蛋白基因表达的影响(英文)[J].中国组织工程研究与临床康复,2007,11(48):9805-9808. 被引量:1

引证文献1

相关作者