期刊文献+

迈瑞BS-200全自动生化分析仪性能评价 被引量:4

Performance Evaluation of Mindray BS-200 Automatic Biochemistry Analyzer
下载PDF
导出
摘要 目的评价深圳迈瑞BS-200全自动生化分析仪的主要性能。方法选取涵盖各波长的8个项目进行测定,对BS-200全自动生化分析仪进行重复性、线性、精密度、准确度进行评价,并与日立7170全自动生化分析仪进行对比试验。结果日间、批内、批间及总变异系数较小,重复性好,准确度高,线性范围较宽。与日立71710的对比试验,相关系(r)均>0.98。结论该仪器的各评价指标均较好,值得在中小医院推广应用。 Objective To evaluate the main performance of BS-200 automatic biochemistry analyzer. Methods To evaluate the repeatability, linearity, precision and accurate Of the BS-200 automatic biochemistry analyzer by 8 items that contain all wavelength(ALT. GGT. Cl-, Cr. Glu. TP. AIb. APOA1. TBil), and carry on the contrast experiment with Hitachi 7170, Results The daytime run CV,within-run CV between-run CV and total CV are under 4.0%, the duplication is well, the accuracy is high, the linear scope is wide, and the contrast experiment correlation is good. Conclusion This instrument each appraisal target is good, and worth in the small or middle-sized hospital promotion application.
出处 《中国医疗器械杂志》 CAS 2009年第3期221-222,共2页 Chinese Journal of Medical Instrumentation
关键词 BS-200全自动生化分析仪 变异系数 准确度 BS-200 Automatic biochemitry analyzer, Coefficient of Variation, Accuracy