摘要
本文根据Datex Ohmeda 7100型麻醉机的基本工作原理,推导出吸入潮气量的计算公式,经临床验证是正确的。并对如何利用公式指导麻醉机和监测仪的临床应用及维修作了概述。
According to the basic work principle of Datex Ohmeda 7100 anesthetic machine, the calculation formula of inspiratory tidal volume is deduced in the paper. The formula is verified by clinical application. The summary how to direct the applying and maintaining anesthetic machine and its monitoring instrument by using formula is presented.
出处
《中国医疗器械杂志》
CAS
2009年第3期223-224,共2页
Chinese Journal of Medical Instrumentation
关键词
麻醉机
吸入潮气量
监测仪
anesthetic machine, inspiratory tidal volume, monitoring instrument