以PMBP为萃取剂和化学改进剂/电热蒸发电感耦合等离子体原子发射光谱测定二氧化锆中痕量铕 被引量:6

Determination of Trace Europium in Zirconium Dioxide by Electrothermal Vaporization Inductively Coupled Plasma -Atomic Emission Spectrometry With 1-Phenyl -3-methyl -4-benzoyl -pyrazone [5] as Extractant and Chemical Modifier
下载PDF
导出
摘要 提出了用β-二酮类试剂1-苯基-3-甲基-4苯甲酰基-吡唑酮[5](PMBP)为萃取剂和 化学改进剂的电热蒸发电感耦合等离子体原子发射光谱(ETV—ICP—AES)测定高纯ZrO2 中痕量Eu的新方法。讨论了影响Eu-PMBP 螯合物蒸发的主要因素。在优化实验条件下,Eu 的检出限为0.9μg/L,相对标准偏差为3.4%,试样分析的回收率为94.5%~105.2%. Based on the formation of volatile metal chelate between metal ion and β-diketone reagent , a new method is reported for the determination of trace europium in high purity zirconium dioxide by electrothermal vaporization-inductively coupled plasma-atomic emission spectrometry (ETV-ICP-AES) with l-phenyl-3-methyl-4-benzoyl (5) (PMBP) as extractant for removal of Zr matrix and chemical modifier for the determination of Eu. The vaporization behavior, effect factors and analytical characteristics of above determined element were investigated in detail. Under optimized experimental conditions, the detection limit, relative standard deviation, and recovery are 0. 9 μg/L, 3. 4% (C=0. 5 mg/L, V = 10 μL, n = 7) and 94. 5%-105. 2%, respectively. The linear range of the proposed method covers three orders of magnitude. The method has been applied for analysis of real sample, and the results are satisfactory.
出处 《分析化学》 SCIE EI CAS CSCD 北大核心 1998年第6期762-765,共4页 Chinese Journal of Analytical Chemistry
关键词 高纯二氧化锆 二氧化锆 测定 分析 PMBP Electrothermal vaporization-Inductively coupled plasma-atomic emission spec- trometry, zirconium dioxide, rare earth element, europium, 1-phenyl-3-methyl-4-benzoyl pyrazone [5]
  • 相关文献

参考文献4

  • 1彭天右,J Anal Atom Spectrom,1998年,13卷,75页
  • 2彭天右,中国稀土学报,1997年,15卷,257页
  • 3Tao S,J Anal Atom Spectrom,1996年,11卷,111页
  • 4Tao S,Anal Proc,1995年,32卷,371页

同被引文献969

引证文献6

二级引证文献34

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部