报道了伽玛治疗刀的辐射屏蔽设计计算方法和计算结果。本伽玛治疗刀由30个每个7.4EBq的60Co点源,半球状屏蔽层、侧屏蔽柜和防护门组成,计算了后屏蔽体,侧屏蔽柜和防护门的屏蔽厚度,对泄漏辐射的监则结果表明,本伽玛治疗刀的屏蔽是足够安全的,计算的空气比释动能率的限制值和测量值相符合。
This paper reportes the method and results of radiation shielding calculation of the gamma knife for therapy which is composed of thirty 60 Co sourses each with 7.4EBq, semi-spherical shield, lateral shielding cupboard and the shielding door. The shielding thicknesses of the back shield, the lateral shielding capboard and the shielding door were calculated. The leakage radiation by test indicates that the shielding is sufficient safety for this Gamma knife and the Kerma rate of control calculated agrees with that by test.
出处
《核科学与工程》
CSCD
北大核心
1998年第2期165-171,共7页
Nuclear Science and Engineering