期刊文献+

微量进样火焰原子吸收法测定植物油中镁和钙 被引量:4

Determination of Mg and Ca in vegetable oils by flame atomic absorption spectrometry using micro amount injection technique
下载PDF
导出
摘要 采用干灰化法处理豆油、棉籽油和芝麻油样品(480℃、5h)时,将常规火焰原子吸收进样方式改成微量注射进样,测定了油中镁、钙含量。每次进样20μL,测镁特征浓度为8.19ng/g,检出限为0.015μg/g,相对标准偏差为2.6%;测钙特征浓度为0.11ng/g,检出限为0.067μg/g,相对标准偏差为3.1%。三种植物油中镁含量分别为20.54、23.92和20.75μg/g,钙含量分别为4.72、2.33和3.00μg/g。 Micro amount injection technique for the determination of Mg and Ca in soyabean, cotton seed and sesame oils by FAAS is described . The micro amount injector is and adapted conventional FAAS sample introduction device. The sample volume of each measurement is 20μL. The characteristic concentration , lower detection limit and RSD for Mg are 8.19ng/g , 0.015μg/g, 2 .6%,respectively, for Ca 0. 11ng/g, 0.067μg/g, 3. 1 %, respectively. The concentrations of Mg in soyabean, cotton seed and sesame oils are 20.54, 23.92 and 20.75μg/g, respectively, that of Ca 4.72, 2.33 and 3.00μg/g, respectively
机构地区 河北大学化学系
出处 《分析仪器》 CAS 1998年第3期33-35,共3页 Analytical Instrumentation
  • 相关文献

参考文献3

  • 1孙汉文,原子吸收光谱分析技术,1992年
  • 2苏文周,分析测试通报,1988年,7卷,4期,39页
  • 3陈树瑜,分析化学,1988年,16卷,15期,460页