摘要
目的:建立一次性使用输液器中环己酮残留量的测定方法。方法:利用蠕动泵对输液器中的环己酮进行提取,采用气相色谱法进行测定。色谱柱为HP-FFAP石英毛细管柱(30m×0.53mm×1.0μm);进样口温度200℃;柱温90℃;氢火焰离子化检测器(FID),检测器温度260℃;载气为氮气;流量1.7mL·min-1;分流比:5∶1;进样体积1.0μL;以无水乙醇为内标物质,以水为溶剂。结果:环己酮与内标物质在4min内分离良好;线性相关系数为0.9997;平均回收率为100.1%(n=9)。结论:该方法简单、快速、准确,可用于一次性使用输液器中环己酮残留量的检测。
Objective: To establish a method for the determination of residual cyclohexanone in disposable infusion set. Methods: Residual cyclohexanone in disposable infusion set was extracted with circular pump, then it was determined by GC with HP - FFAP capillary column, FID detector and using nitrogen as carrier gas. The injection temperature was 200 ℃, the column temperature was 90 ~C and the detector temperature was 260 ℃. The flow rate of nitrogen was 1.7 mL · min^-1. The split ratio was 5: 1. The injection volume was 1.0 μL. The internal standard was alcohol , and the solvent was water. Results:The cyclohexanone and internal standard were separated perfectly with 4 min. The correlation coefficient of the linear equation was O. 9997, the average recovery was 100. 1% ( n = 9). Conclusion: The method is simple, rapid and accurate for the determination of residual cyclohexanone in disposable infusion set.