加速器机架角度对多叶准直器叶片到位精度的影响 被引量:14

Impact of accelerator's gantry angle on multi-leaf collimator position accuracy
原文传递
导出
摘要 目的测量加速器机架角度对多叶准直器叶片到位精度的影响。方法用柯达xomat—V胶片进行测量,选定合适叶片位置和机架角度,保证精确摆位,用RITll3胶片分析软件对曝光的胶片做分析。结果在所有测量情况下,绝大部分叶片到位偏差都〈0.5mm,只有在机架角270°、叶片从左向右运动情况下有2对叶片到位偏差〉1mm。结论机架角度即重力对多叶准直器叶片到位精度确有影响,精确摆位对测量工作至关重要。 Objective To study the impact of accelerator's gantry angle on MLC position accuracy. Methods Kodak X-omat-V Films were used, appropriate multi-leaf collimator positions and gantry angles were chosen , precise setup was needed , and the exposal films were analyzed by RIT 1 1 3 software. Results At all conditions, most of the leaf position errors were less than 0.5 ram. There were 2 pairs of leaf position errors larger than 1 mm when the gantry angle was 270° and the leaves moved from left to right. Conclu- sions Gantry angle can affect the accuracy of multi-leaf collimator position. Gravity and inertia increase the uncertainty of leaf position accuracy. Pricise setup is critical for the dose measurement.
出处 《中华放射肿瘤学杂志》 CSCD 北大核心 2009年第4期317-320,共4页 Chinese Journal of Radiation Oncology
关键词 调强放射疗法 机架角度 多叶准直器 到位精度 Intensity modulated radiotherapy Gantry angle Multi-leaf collimator Positionaccuracy
  • 相关文献

参考文献7

二级参考文献12

  • 1马金利,蒋国梁,傅小龙,廖源,吴开良,周莉钧.窄带野胶片检查在多叶准直器质量保证中作用探讨[J].中华放射肿瘤学杂志,2004,13(2):123-127. 被引量:10
  • 2马金利,蒋国梁,傅小龙,李龙根.冲洗时间因素对胶片剂量的影响[J].中华放射肿瘤学杂志,2005,14(6):514-518. 被引量:6
  • 3孙朝阳,肖锋,石梅,魏丽春,陆军.Varian多叶准直器(MLC)常规测量方法及质量保证[J].中国医学物理学杂志,2005,22(6):697-698. 被引量:10
  • 4Lossasso T, Chui CS, Ling CC. Comprehensive quality assurance for the delivery of intensity modulated radiotherapy with a multileaf collimator used in the dynamic mode. Med Phys, 2001, 28:2209-2219.
  • 5Essers M, de Langen M, Dirkx MLP, et al. Commissioning of a commercially available system for intensity-modulated radiotherapy dose delivery with dynamic multileaf collimation. Radiother Oncol, 2001,60:215-224.
  • 6Budgell GJ, Mort JH, Williams PC, et al. Requirements for leaf position accuracy for dynamic multileaf collimation. Phys Med Biol, 2000, 45:1211-1227.
  • 7Saw CB, Ayyangar KM, Thompson RB, et al. Commissioning and quality assurace for MLC-based IMRT. Med Dosim, 2001, 26:125-133.
  • 8Mubata CD, Childs P, Bidmead AM. A quality assurance procedure for the Varian multi-leaf collimator[J]. Phys. Med. Biol, 1997,42:423-431.
  • 9Childress NL, Dong L. Rapid radiographic film calibration for IMRT verification using automated MLC field. Med Phys,2002,47,4121- 4130.
  • 10Ravi K. Accuracy of rapid radiographic film calibration for intensitymodulated radiation therapy verification. J Appl Clin Med Phys. , 2006 May 25 ;7(2) :86-95.

共引文献20

同被引文献79

引证文献14

二级引证文献66

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部