Topslane体部定位框架患者定位准确性分析 被引量:3

Evaluation of setup accuracy for Topslane body stereostatic frame
下载PDF
导出
摘要 目的:分析Topslane体部定位框架的定位准确性。方法:对31例胸部肿瘤患者应用Topslane体部定位框架进行定位,患者体表放置定位标记,CT扫描后经DICOM 3.0传输到治疗计划系统,分析体表参考标记物计算值和测量值的坐标偏差,确定体部定位框架的定位准确性。结果:Topslane体部定位框架定位准确性误差在±2 mm之内。结论:Topslane体部定位框架具有良好的机械性能,操作安全简便,是适形放疗中患者靶区定位和位置验证的辅助设备。 Objective: To evaluate the positioning accuracy of the Topslane body localization frame.Methods: 31 cases of chest tumor were immobilized using the Topslane body localization frame.Fiducial markers were deliberately placed on each patient's skin. CT images were transferred to TPS via DICOM 3.0.Calculated and measured coordinates of markers were compared.Results: The positioning accuracy of the Topslane body localization was found to be within±2 mm. Conclusion: The Topslane body localization frame has good mechanical property.It is simple for use in patient immobilization and setup verification.
出处 《医疗装备》 2009年第8期49-51,共3页 Medical Equipment
关键词 体部定位框架 定位准确性 适形放疗 body localization frame setup accuracy Conformal radiotherapy
  • 相关文献

参考文献3

  • 1International Organization for Standardization."Quality Management and Quality Assurance Standards-Part 1:Guideliness for Selection and Use."ISO 900O,1994.
  • 2Kutcher,G.J.,L.Coia,M.Gillin,W.F.Hanson,S.Leibel,R.J.Morton,J.R.Palta,J.A.Purdy,L.E.Reinstein,G.K.Svensson,M.Weller,L.Wingfield,"Comprehensive QA for radiation oncology."Report of AAPM Radiation Therapy Committee Task Group 40."Med.Phys.21:581-618.(1994).
  • 3Purdy,J.A.,and C.A.Perez,"Quality assurance in radiation oncology in the United States."Rays 21:505-540(1996).

同被引文献20

  • 1Hunt M.The effect of positional uncertainties on the treatment of primary nasopharynx cancer[J].Med Phys,1989,16(3):46.
  • 2Stroom JC,Heijmen BJ.Geometrical uncertainties, radiotherapy planning margins and the ICRU-62 report[J].Radiother Oncol,2002,64(1):75- 83.
  • 3t3ijhold J,van Herk M,VijlbriefR,et al.Fast evaluation of patient set-up during radiotherapy by aligning features in portal and simulator imagesO].Phys Med Bio1,1991,36(12):1665-1679.
  • 4Hurkmans CW,Remeijer P,Lebesque JV,et al.Set-up verification using portal imaging; review of current clinical practice[J]. Radiother 0ncol,2001,58(2):105-120.
  • 5Hunt M.The effect of positional uncertainties on the treatment of primary nasopharynx cancer[J].Med Phys,1989,16(3):46.
  • 6Stroom JC,Heijmen BJ.Geometrical uncertainties, radiotherapy planning margins and the IC1KU-62 report[J].Radiother 0ncol,2002,64(1):75-83.
  • 7Bijhold J,van Herk M,VijlbriefR,et al.Fast evaluation of patient set-up during radiotherapy by aligning features in portal and simulator images[j].Phys Med Bio1,1991,36(12):1665-1679.
  • 8Hurkmans CW,Remeijer P,Lebesque JV,et al.Set-up verification using portal imaging; review of current clinical practice[J]. Radiother Oncol,2001,58(2):105-120.
  • 9肖锋,孙朝阳,胡明民,陆宙,刘晓莉,陆军,石梅.三维适形及调强放疗摆位误差分析[J].中国医学物理学杂志,2008,25(3):641-642. 被引量:21
  • 10曾自力.三维适形放射治疗计划设计的探讨[J].中国辐射卫生,2008,17(2):202-203. 被引量:1

引证文献3

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部