期刊文献+

串联质谱仪不同扫描方式对检测美国CDC酰基肉碱定值质控纸片检测结果的影响 被引量:6

Effects of Scan Types of the Tandem Spectrum to Detect Acylca- rnitines on the Quantitative Dry Blood Spots of CDC in USA
下载PDF
导出
摘要 目的通过探讨串联质谱仪不同扫描方式对检测美国CDC酰基肉碱定值质控纸片结果的影响,寻找检测酰基内碱的最适扫描方式。方法分别采用母离子扫描法(precursorion,Prec)和多反应监测法(multiplereactionmonitoring,MRM)对美国CDC提供的酰基肉碱定值纸片(包括4个浓度水平)进行检测。每天测定所有水平的酰基内碱,连续测定5d。计算不同扫描方式测得的酰基肉碱的批内、批间CV%和相对偏倚。结果Prec法检测C2的批内CV%高于MRM法检测的批内C-W%(f=5.053,P=0.002);Prec法检测C6的批间CV%高于MRM法检测的批阃CV%0=2.737,P=0.034);Prec法和MRM法的相对偏倚均在美国CDC要求的范围之内。结论与Prec法比较,MRM法在检测个别酰基内碱时的精密度更好,但同时,优化MRM法也比较费时、繁琐。因此,建议将两种方法搭配使用,从而达到结果准确、省时省力、节约成本的综合效果。 Objective To find the best scan type of detecting acylcarnitines by analyzing the results of the quantitative dry blood spots of CDC in USA and exploring the influences. Methods All levels of acylcarnitines in the quantitative dry blood spots of CDC in USA were detected by precursor ion (Prec) and multiple reaction monitoring (MRM). There were four levels of the acylcarnitines. Detected all levels of acylcarnitines every day and there were five continuous days for de- tection. Calculated and analyzed the CV~ in the same day and in different days and bias of the results of acylcarnitines of different scan types. Results There were higher CV% in the same day of C2 detected by Pree than by MRM(t=5. 053,P=0. 002). There were higher CV% in the different days of C6 detected by Prec than by MRM (t = 2. 737,P= 0. 034). It could achieve the requirement of the CDC by the two scan types. Conclusion It is more precise for acylcarnitines to be de- tected by MRM than by Prec. But it need more time and more money for optimization of MRM. It is necessary to combine the two scan types for accurate results ,saving money and time.
出处 《现代检验医学杂志》 CAS 2009年第4期90-93,共4页 Journal of Modern Laboratory Medicine
关键词 串联质谱仪 酰基肉碱 扫描方式 tandem spectrum aeylearnitine scan type
  • 相关文献

参考文献5

  • 1Schulze A,Lindner M,Kohlmüller D,et al.Expanded newborn screening for inborn errors of metabolism by electrospray ionization-tandem mass spectrometry:results,outcome,and implications[J].Pediatrics,2003,111(6Pt1):1399-1406.
  • 2Chace DH,DiPerna JC,Mitchell BL,et al.Electrospray tandem mass spectrometry for analysis of acylcarnitines in dried postmortem blood specimens collected at autopsy from infants with unexplained cause of death[J].Clin Chem,2001,47(7):1166-1182.
  • 3韩连书,高晓岚,叶军,邱文娟,顾学范.串联质谱分析干血滤纸片酰基肉碱方法的建立[J].中华检验医学杂志,2005,28(1):88-91. 被引量:31
  • 4Zytkovicz TH,Fitzgerald EF,Marsden D,et al.Tandem mass spectrometry analysis for amino,organic,and fatty acid disorders in newborn dried blood spots:a two-year summary form the New England newborn screening program[J].Clin Chem,2001,47(7):1945-1955.
  • 5Paglia G,D'Apolito O,Corso G,et al.Precursor ion scan profiles of acylcarnitines by atmospheric pressure thermal desorption chemical ionization tandem mass spectrometry[J].Rapid Commun Mass Spectrom,2008,22(23):3809-3815.

二级参考文献12

  • 1Schulze A, Lindner M, Kohlmuler D, et al. Expanded newborn screening for inborn errors of metabolism by electrospray ionization-tandem mass spectrometry. Pediatrics, 2003, 111:1399-1406.
  • 2Chace DH, DiPerna JC, Mitchell BL, et al. Electrospray tandem mass spectrometry for analysis of acylcarnitines in dried postmortem blood specimens collected at autopsy from infants with unexplained cause of death. Clin Chem, 2001, 47:1166-1182.
  • 3Zytkovicz TH, Fitzgerald EF, Marsden D, et al. Tandem mass spectrometric analysis for amino, organic, and fatty acid disorders in newborn dried blood spots: a two-year summary from the New England Newborn Screening Program. Clin Chem, 2001, 47:1945-1955.
  • 4Chace DH, Kalas TA, Naylor EW, et al. The application of tandem mass spectrometry to neonatal screening for inherited disorders of intermediary metabolism. Annu Rev Genomics Hum Genet, 2002, 3:17- 45.
  • 5Rashed MS, Ozand PT, Bucknall MP, et al.Diagnosis of inborn errors of metabolism from blood spots by acylcarnitines and amino acids profiling using automated electrospray tandem mass spectrometry. Pediatr Res, 1995, 38:324-331.
  • 6Chace DH, Hillman SL, van Hove JL, et al. Rapid diagnosis of MCAD deficiency: quantitatively analysis of octanoylcarnitine and other acylcarnitines in newborn blood spots by tandem mass spectrometry. Clin Chem, 1997, 43:2106-2113.
  • 7Chace DH, DiPerna JC, Kalas TA, et al. Rapid diagnosis of methylmalonic and propionic acidemias: quantitative tandem mass spectrometric analysis of propionylcarnitine in filter-paper blood specimens obtained from newborns. Clin Chem, 2001,47: 2040-2044.
  • 8Schulze A, Lindner M, Kohlmuller D, et al. Expanded newborn screening for inborn errors of metabolism by electrospray ionization-tandem mass spectrometry: results, outcome, and implications. Pediatrics, 2003, 111(6 Pt 1):1399-1406.
  • 9Wilcken B, Wiley V, Hammond J, et al. Screening newborns for inborn errors of metabolism by tandem mass spectrometry. N Engl J Med, 2003,348:2304-2312.
  • 10Insinga RP, Laessig RH, Hoffman GL. Newborn screening with tandem mass spectrometry: examining its cost-effectiveness in the Wisconsin Newborn Screening Panel. J Pediatr, 2002,141:524-531.

共引文献30

同被引文献63

引证文献6

二级引证文献20

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部