期刊文献+

血糖仪偏差图设计与临床应用 被引量:1

Bias Plots of Glucose Meter Systems And Clinical Applications
下载PDF
导出
摘要 目的探讨血糖仪准确性评价方法。方法以血糖仪和参比法结果之差绘制血糖仪偏差图,判断血糖仪的准确性。结果依据IS015197准则,当血糖≤75mg/dI(4.2mmol/L)时,三种血糖仪符合率分别为73.53%,97.06%和100%,〉75mg/dl(4.2mmol/L),符合率分别为89.16%,99.40%,99.40%。结论评价血糖仪准确性应当遵循ISO15197准则,“血糖仪偏差图”能够直观地显示血糖仪结果的准确性。 Objective To study glucose meter systems accuracy. Methods "Bias plots" were done by differeces value of all paired glucose meter systems and reference measurements and meters were evaluted. Results Depending on ISO15197 guideline,according rates of three meters were 73. 53%,97.06% and 100% respectively when results ≤75 mg/d,while〉 75 mg/dl were 89. 16% ,99.40% and 99.40%. Conclusion Evaluation of meters must follow ISO15197 guideline. Accuracy of systems can be judged by "bias plots"visually.
出处 《现代检验医学杂志》 CAS 2009年第4期123-124,共2页 Journal of Modern Laboratory Medicine
关键词 血糖仪偏差图dS015197准则 bias plots ISO15197 guiedline
  • 相关文献

参考文献1

  • 1国家食品药品监督管理局.GB/T19634-2005体外诊断检验系统自测用血糖监测系统通用技术条件[S].2006:7.

共引文献2

同被引文献14

引证文献1

二级引证文献10

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...