摘要
目的:评价Sysmex XE-2100血细胞分析仪检测外周血标本时有核红细胞(NRBC)异常提示Q-flag值的可靠性,并探讨其临床应用价值以及影响因素。方法:将NRBC提示阳性(Q-flag值100-300)的血标本100例及无NRBC提示异常(Q-flag值100以下)的血标本100例进行手工涂片瑞特染色镜检,以显微镜检结果为金标准,判断Q-flag值提示的可靠性。结果:SysmexXE-2100测定NRBC提示的灵敏度、特异性分别为98.7%、80.5%。结论:Q-flag值在有核红细胞的检测中有一定的应用意义。
Objective:To evaluate reliability of the NRBC hinting function (Q-flag) of Sysmex XE-2100 automatic hematology analyzer and investigate its clinical application and influencing factors. Method.. 100 samples with abnormal NRBC(Q-flag value〉100) hinted by the hematology analyzer and 100 samples without NRBC hints were detected with microscopy,and result of manual microscopic examination was used as a golden standard to judge the sensitivity and specificity of the hematology analyzer. Result:The sensitivity and specificity of hematology analyzer Q-flag was 95.9% and 81%, respectively. Concluslon..The Q-flag value had a certain application significance in NRBC detection
出处
《临床血液学杂志(输血与检验)》
CAS
2009年第4期423-424,432,共3页
Journal of Clinical Hematology(Blood Transfusion & Laboratory Medicine)
关键词
有核红细胞
血细胞分析仪
血细胞计数
显微镜镜检
nucleated red blood cell
hematology analyzer
blood cell count
microscopic examination