螺旋断层放疗MVCT成像剂量测量及其分析 被引量:13

Measurement and analysis of the imaging dose with megavoltage computed tomography for helical tomotherapy
下载PDF
导出
摘要 背景与目的:螺旋断层放疗中治疗前/后通常采用MVCT(megavoltage computed tomography,MVCT)来实行图像引导下精确定位,成像过程中势必会增加患者的辐射剂量。通过体部及头部标准模体分别模拟患者MVCT成像条件来实现其剂量学测量及研究。方法:采用螺旋断层治疗机MVCT分别对圆柱形体部模体和自制标准头模体在不同螺距比(1.0、2.0、3.0)、扫描长度(4.8、7.2、9.6、12、14.4cm)及摆位条件下进行扫描,测量模体内各点扫描平均剂量变化,并对其剂量均匀性情况进行定性分析。结果:体部模体内MVCT扫描辐射剂量在0.599~2.876cGy之间;而头部模体点平均剂量为0.913~3.231cGy。测量结果表明,模体内所受照射剂量与螺距比值、扫描长度呈密切相关性;且MVCT扫描剂量与CT螺距值基本成反比关系。不同摆位条件下辐射剂量值沿模体径向大小而相应产生出其变化趋势。结论:实验结果有助于为治疗计划设计及图像引导治疗扫描范围提供较为现实而可靠的临床依据,减小患者治疗期间可能诱发肿瘤所获取的额外剂量。 Background & Objective: During the helical tomotherapy process, megavoltage computed tomography (MVCT) images are usually used for guiding the precise setup of patients before/after treatment delivery, which would certainly increase the total dose for patients. This study was to investigate the imaging dose of MVCT using the body and head phantom on a tomotherapy machine. Methods: A set of cylindrical body and head phantoms was adopted for scanning with different pitch values (1.0/2.0/ 3.0), lengths (4.8/7.2/9.6/12/14.4 cm) and patient setups on the couch of tomotherapy system. The average MVCT imaging doses were measured using AlSL chambers inserted in the phantoms with preset geometry. The dose uniformity was qualitatively analyzed. Results:The MVCT scanning dose for the body phantom was between 0.599 and 2.876 cGy during each treatment delivery, while the dose for the head phantom was between 0.913 and 3231 cGy. Two major parameters, the assigned pitch numbers and scanning lengths, were the most important impacts to the dose variation. The MVCT dose was inversely proportional to the CT pitch value. With respect to the scanning length, the doses responded differently along the radial direction of the phantoms with different setup criteria. Conclusion. The results may provide a reliable guidance for proper planning design of the scanning region, which is valuable to help minimize the extra doses to patient.
出处 《癌症》 SCIE CAS CSCD 北大核心 2009年第8期886-889,共4页 Chinese Journal of Cancer
关键词 螺旋断层 放射治疗 MVCT 成像剂量 helical tomotherapy, radiotherapy, MVCT, imaging dose
  • 相关文献

参考文献8

  • 1Meeks SL, Harmon J J, Langen KM, et al. Performance characterization of megavoltage computed tomography imaging on a helical tomotherapy unit [J]. Med Phys, 2005,32(8):2673 - 2681.
  • 2Mackie TR, Balog J, Ruchala K, et al. Tomotherapy [J]. Semin Radiat Oncol, 1999,9( 1 ) : 108-117.
  • 3Ruchala K J, Olivera GH, Schloesser EA, et al. Megavoltage CT on a tomotherapy system [J]. Phys Med Biol, 1999,44 (10) : 2597-2621.
  • 4Mutic S, Palta JR, Butker EK, et al. Quality assurance for computedtomography simulators and the computetomography-simulation process : report of the AAPM Radiation Therapy Committee Task Group No.66 [J]. Med Phys, 2003,30(10) : 2762-2792.
  • 5Shah AP, Langen KM, Ruchala K J, et al. Patient dose from megavoltage computed tomography imaging [J]. Int. J. Radiat Oncol Biol Phys, 2008,70 (5) : 1579-1587.
  • 6刘彬,白玫,费晓璐,魏岚.CT剂量表征量的选择[J].中国医疗设备,2008,23(3):54-56. 被引量:9
  • 7Zeidan OA, Langen KM, Meeks SL, et al. Evaluation of image-guidance protocols in the treatment of head and neck cancers [J]. Int J Radiat Oncol Biol Phys, 2007,67 (3) : 670-677.
  • 8Beavis AW. Is tomotherapy the future of IMRT [J]? Br J Radio, 2004, 77(916) : 285-295.

二级参考文献8

  • 1刘秀珍.X-CT辐射剂量的测量[J].医疗卫生装备,2006,27(2):61-62. 被引量:5
  • 2白玫.浅谈螺旋CT辐射剂量的测量[J].医疗设备信息,2006,21(7):33-35. 被引量:5
  • 3[2]GB 18871-2002,电离辐射防护与辐射源安全基本标准[S].
  • 4[1]United States Food and Drug Administration Center for Devices and Radiological Health.Whole boay scanning:using computed tomography(CT)[S/OL].Available at:http://www.fda.gov/cdrh/ct.
  • 5[2]European Guidelines on Quality Criteria for Computed Tomography[S/OL].Available at:http://www.dr s.dk/ guiddines/ct/q uality/index.htm.
  • 6[5]International Commission on Radiological Protection,Publication 60:1990 Recommendations of the International ommission on Radiological Protection[S].Annals of the ICRE Vol 21,Nos.13 (Pergamon,Oxford,1991).
  • 7[6]XU X G,CHAO T C,and BOZKURT A.VIP-Man:An image-based whole-boay adult male model constructed from color photographs of the visible human project for multi-particle[J].Monte Carlo calculafiom Health Physics,78(5)476-486,(2000).
  • 8吴毅,杜国生,田中青.X射线计算机断层摄影(CT)受检者剂量检测方法的研究[J].辐射防护,1998,18(3):185-190. 被引量:8

共引文献8

同被引文献166

引证文献13

二级引证文献61

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部