期刊文献+

基于线性调频信号的高帧率超声成像系统的改进 被引量:1

Improvement on High Frame Rate Ultrasonic Imaging System Based on Linear Frequency-modulated Signal
原文传递
导出
摘要 线性调频信号用于高帧率超声成像系统能实现快速成像,并能显著提高成像系统的信噪比,但回波信号通过匹配滤波器后,输出压缩脉冲的旁瓣会引起较小目标的丢失。我们对基于线性调频信号的高帧率超声成像系统进行改进。采用预失真的线性调频信号作为激励信号,在接收端对接收信号通过失配滤波器处理。结果表明基于线性调频信号的高帧率超声成像系统经过改进之后,在提高系统信噪比的同时也能提高轴向分辨率。 The high frame rate (HFR) ultrasonic imaging system based on linear frequency-modulated (LFM) signal constructs images at a high frame rate; the signal-to-noise ratio (SNR) of this system can also be improved. Unfortunately, such pulse compression methods that increase the SNR usually cause range sidelobe artifacts. In an imaging situation, the effects of the sidelobes extending on either side of the compressed pulse will be self-noise along the axial direction and masking of weaker echoes. The improvement on high frame rate ultrasonic imaging system based on LFM signal is considered in this paper. In this proposed scheme, a predistorted LFM signal is used as excited signal and a mismatched filter is applied on receiving end. The results show that the proposed HFR ultrasonic imaging system can achieve higher SNR and the axial resolution is also improved.
出处 《生物医学工程学杂志》 EI CAS CSCD 北大核心 2009年第4期761-765,共5页 Journal of Biomedical Engineering
关键词 高帧率 预失真线性调频信号 失配滤波 High frame rate(HFR) Predistorted linear frequency-modulated signal Mismatched filter
  • 相关文献

参考文献8

二级参考文献24

  • 1彭虎 ,Lu Jianyu ,冯焕清 ,张作生 .基于角谱原理的高帧率超声成像系统的计算机仿真[J].生物医学工程学杂志,2004,21(5):766-769. 被引量:6
  • 2Jianyu Lu and Greenleaf J F. Nondiffraeting X waves-exact solutions to free-space scalar wave equation and their finite aperture realizations. IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1992, 39 (1): 19-31.
  • 3Jianyu Lu and Greenleaf J F. Experimental verification of nondiffracting X waves. IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1992, 39 (3): 441-446.
  • 4Jian yu Lu. 2D and 3D high frame rate imaging with limited diffraction beams. IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1997, 44 (4): 839-856.
  • 5Jian yu Lu. Experimental study of high frame rate imaging with limited diffraction beams. IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1998, 45 (1): 84-97.
  • 6Matthew O' Donnell. Coded excitation system for improving the penetration of real-time phased-array imaging system. IEEE Transactions of Ultrasonics, Ferroelectrics, and Frequency Control, 1992, 39 (3):341 - 351.
  • 7Goodman J W. Introduction to fourier optics. New York: McGraw-Hill,1968, chs. 2-4.
  • 8Lu Jian-yu and Greenleaf J F. Nondiffracting X waves - exact solutions to free - space scalar wave equation and their finite aperture realizations [ J ]. IEEE Transactions on Ultrasonics,Fermelectrics, and Frequency Control, 1992, 39(1) : 19-31.
  • 9Lu Jian-yu and Greenleaf J F. Experimental verification of nondiffracting X waves [ J ].IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1992, 39 ( 3 ) :441-446.
  • 10Peng Hu, Zhang Zuo-sheng, Lu Jian-yu. A study of simulation method of limited diffraction beams[ J ]. The Journal of the Acoustical Society of America, 2000, 108 (5) : 2595-2595.

共引文献23

同被引文献18

引证文献1

二级引证文献8

相关作者