纯铁在几种不同介质中的腐蚀行为

The Corrosion of Pure Iron in Five Different Mediums
原文传递
导出
摘要 采用计划化的间歇腐蚀试验方法(分段试验法)研究了纯铁在0.15mol/LNaCl溶液,Ringer’s溶液,PBS(-)溶液,SBF溶液以及M199培养基中,浸泡不同时间材料腐蚀性能。结果表明,不同模拟体液对铁的侵蚀性是不同的,侵蚀性由强到弱的顺序为0.15mol/LNaCl(Ringer’s)>PBS(-)>SBF>M199培养基,SEM及EDX分析显示在模拟体液中添加HPO42-、H2PO4-、Ca2+、Mg2+、SO42-以及氨基酸等有机成分起到了一定的缓蚀作用。 The sectional test was adopted in this study to investigate the corrosion of pure iron in 0. 15 mol/L NaCl solution, Ringer solution, PBS(-) solution, SBF solution and M199 cell culture medium at three different times. The result shows that different mediums have different corrosion effects on pure iron. The arrangement according to the medium's corrosion ability from the strongest to weakest is 0.15 mol/L NaCl solution (Ringer solution) ,PBS(-) solution, SBF solution and M199 cell culture medium. The results of scanning electron microscopy and energy dispersive X-ray spectrum analyses show that the addition of HPO4^2- , H2 PO4^- , Ca^2+ , Mg^2+ , SO4^2- and the organic component can inhibit the corrosion to some degree.
出处 《生物医学工程学杂志》 EI CAS CSCD 北大核心 2009年第4期783-786,共4页 Journal of Biomedical Engineering
基金 国家自然科学基金资助项目(30670565)
关键词 纯铁 模拟体液 腐蚀 Iron Simulate body fluid(SBF) Corrosion
  • 相关文献

参考文献6

  • 1潘长江,王进,黄楠.血管支架内再狭窄的研究进展[J].中国生物医学工程学报,2004,23(2):152-156. 被引量:26
  • 2赵振心,刘道志,张一.血管支架材料及其临床研究进展[J].中国医疗器械杂志,2005,29(6):391-395. 被引量:16
  • 3MATTHIAS PEUSTERA, CAROLA HESSEA, TIRZA SCHLOOA,et al. Long-term biocompatibility of a corrodible peripheral iron stent in the porcine descending aorta[J]. Biomaterlals, 2006,27: 4955-4962.
  • 4中华人民共和国国家标准GB10124-88,金属材料实验室均匀腐蚀全浸试验方法,附录A[S].
  • 5HANAWA T,OTA M. Characterization of surface film formed on titanium in electrolyte using XPS[J]. Applied Surface Science,1992,55:269.
  • 6JEYAPRABHA C, SATHIYANARAYANAN S, VENKATACHARI G. Corrosion inhibition of pure iron in 0.5 M H2SO4 solutions by ethanolamines[J]. Applied Surface Science, 2005, 246:108-116.

二级参考文献36

  • 1Dotter CT, Judkins MD. Transluminal treatment of arteriosclerotic obatruction: a new technique and a preliminary report of its application[J]. Circulation,1964,30:654.
  • 2Dotter CT. Transluminally placed coil spring endoarterial tube grafts, long term patency in canine politeal artery[J]. Invest Radiol,1969,4:327.
  • 3Baurschmidt P, Schaldach M. Electrochemical aspects of the thrombogenicity of a materials[J]. J Bioengng,1997,1(11):261.
  • 4Sigwart U, Puel J, Mirkovitch V, et al. Intravascular stents to prevent occlusion and restenosis after transluminal grafting[J]. Radiology,1986,160:723.
  • 5Gunn J, Malik N, Shepherd L, et al. In-stent restenosis: more metal and more symmetry required[J]. Heart,1997,77:46.
  • 6E Regar, G Sianos, PW Serruys. Stent development and local drug delivery[J]. British Medical Bulletion,2001,59:227.
  • 7Mehran R, Dangas G, Abizaid A, et al. Angiographic patterns of in-stent restenosis: classification and implications for long-term outcome[J]. Circulation,1999,100:1872.
  • 8Mach F. Toward new therapeutic strategies against neointimal formation in restenosis. Atheroscler Thromb Vasc Biol,2000,20:1699.
  • 9Mudra H, Regar E, Klauss V, et al. Serial follow-up after optimized ultrasound guided deployment of Palmaz-Schatz stents[J]. Circulation,1997,95:363.
  • 10Hoffman R, Minta GS, Dussaillant RG, et al. Patterns and mechanisms of in-stent restenosis: a serial intravascular ultrasound study[J]. Circulation,1996,94:1247.

共引文献39

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部