基于投影寻踪的超声彩色血流成像杂波抑制方法

A clutter rejection method based on the projectionpursuit in the ultrasound color flow imaging
下载PDF
导出
摘要 为抑制超声彩色血流成像(CFI)中杂波信号对血流速度估计的影响,提出一种基于投影寻踪(PP)的杂波抑制方法。根据回波信号的时域特性,采用投影寻踪法提取主元以降低主元对非血流区域的高强度杂波的敏感度;然后在特征向量空间去除估计得到的杂波成分,从而获得血流信号。将本方法用于计算机仿真信号和人体实测信号。结果表明:与传统的高通滤波器法相比,本方法能更好地保持流速剖面的完整性和血流轮廓;与特征向量法相比,在近似的流速完整度下本方法在非血流区域产生的误差较少,血管内外能量比高出特征向量法约5dB,能够有效地提高彩色血流成像的质量。 In the ultrasound color flow imaging (CFI), a clutter rejection method based on the projection-pursuit (PP) is proposed to reduce velocity artifacts caused by clutter signals. According to time domain characteristics of echoes, principal components are extracted by the PP to reduce their sensitivity to strong clutter signals originating from non- blood-flow regions. Blood signals are then obtained by removing the estimated clutter space in the eigenvector domain. The proposed method has been applied to both computer simulated data and human experimental data. Results show that this method can achieve a better velocity profile and vessel shape than the conventional high-pass filter. It is also shown that this method causes less velocity artifacts in non-blood-flow regions and achieves an inside to outside blood energy ratio higher than that achieved by eigenvector method by 5 dB. Thus, the flow image quality is expected to be improved effectively.
作者 尤伟 汪源源
出处 《声学学报》 EI CSCD 北大核心 2009年第5期430-436,共7页 Acta Acustica
基金 国家重点琏础研究规划基金(2005CB724303) 上海市重点学科建设资助项目(B112)
关键词 杂波信号 投影寻踪 抑制方法 血流 成像 彩色 超声 特征向量法 Acoustic waves Blood Clutter (information theory) Eigenvalues and eigenfunctions High pass filters Image quality Pumping plants Ultrasonics
  • 相关文献

参考文献17

  • 1Bjaerum S, Torp H, Kristoffersen K. Clutter filters adapted to tissue motion in ultrasound color flow imaging. IEEE Trans. Ultrason., Ferroelect., Freq. Contr., 2002; 49(6): 693-704.
  • 2陶倩,汪源源,Cardoso Jose,王威琪.管壁和血流多普勒信号的分离研究[J].声学学报,2004,29(4):329-333. 被引量:7
  • 3金大伟,汪源源,王威琪.空间选择性降噪法提取血流超声多普勒信号[J].声学学报,2007,32(3):245-249. 被引量:8
  • 4Chornoboy E S. Initialization for improved IIR filter performance. IEEE Trans. Signal Processing, 1992;40(3): 543--550.
  • 5Kadi A P, Loupas T. On the performance of regression and step-initialized IIR clutter filters for color Doppler systems in diagnostic medical ultrasound. IEEE Trans. Ultrason., Ferroelect., Freq. Contr., 1995; 42(9): 927 937.
  • 6Bjaerum S, Torp H, Kristoffersen K. Clutter filter design for ultrasound color flow imaging. IEEE Trans. Ultrason., Ferroelect., Freq. Contr., 2002; 49(2): 204--216.
  • 7Ledoux L A F, Brands P J, Hoeks A P G. Reduction of the clutter component in Doppler ultrasound signals based on singular value decomposition: A simulation study. Ultrason. Imag., 1997; 19:1--18.
  • 8Bjaerum S, Torp H. Optimal adaptive clutter filtering in color flow imaging. Proc. IEEE Ultrason. Symp., 1997; 2: 1223--1226.
  • 9Yu A C H, Cobbold R S C. Single-ensemble-based eigenprocessing methods for color flow imaging-part I. the Hankel-SVD filter. IEEE Trans. Ultrason., Ferroelect., Freq. Contr., 2008; 55(3): 559--572.
  • 10Li P, Yang X, Zhang D et al. Adaptive clutter filtering based on sparse component analysis in ultrasound color flow imaging. IEEE Trans. Ultrason., Ferroelect., Freq.Contr., 2008; 55(7): 1582--1596.

二级参考文献20

共引文献14

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部