摘要 目的进行离子色谱法测定水中溴酸根离子含量的不确定度分析。方法按照测定程序,建立数学模型,根据不确定度传播规律,计算出相对和绝对扩展不确定度。结果水样中溴酸根离子的浓度为0.0638mg/L时,其绝对扩展不确定度为±0.0031mg/L。结论该方法在实际工作中有较强的应用价值。 [ Objective] To analyze uncertainty of Bromate ion determination in water by ion chromatography. [ Methods ] According to procedures, mathematic model was established , then the relative and the absolute expanded uncertainties were calculated based on the spreading regulation of uncertainty. [ Results] When the concentration of Bromate in water arrived 0.0638 mg/L, the expand- ed uncertainty was ±0. 0031 mg/L. [ Conclusion]This method can produce large application value in practical work.
作者 周小新 周虹
出处 《职业与健康》 CAS 2009年第17期1827-1829,共3页 Occupation and Health
关键词 离子色谱法 溴酸根离子 不确定度 Ion chromatography Bromate ion Uncertainty