HACH DR/2500分光光度法测定水样中总磷的可行性研究 被引量:8

The Feasibility Study of the Mensuration of Total Phosphorus Concentration in Water Sample by HACH DR/2500 Spectrophotometer
下载PDF
导出
摘要 文章运用HACH DR/2500分光光度计,对HACH法测定水样中总磷的检出限、精密度、回收率等一系列实验室质量要求进行了测定。并在此基础上,对多个断面的地表水和不同行业工业废水水体的实际样品进行了HACH法与国家标准钼酸铵分光光度法比对实验,从而证明HACH DR/2500分光光度法与国家标准方法无显著性差异。 In this study, the detection limits, precision and recovery rate as well as other laboratory quality indexes were mensurated in the mensuration of total phosphorus concentration in water sample by HACH DR/2500 speetrophotometcr. Based on experimen, adopting surface water samples collected from several sections and wastewater samples came from different Industries, the comparative experiment has been carried on to detect the outcome discrepancy of total phosphorus concentration between the standard ammonium - molybdate spectrophotometric method and that method by HACH DR/2500 spectrophotometer. It was testified that there was no significant discrepancy between the two methods.
出处 《环境科学与管理》 CAS 2009年第8期125-127,共3页 Environmental Science and Management
关键词 总磷 分光光度法 质量控制 t检验法 total phosphorus spectrophotometry quality control t - test
  • 相关文献

参考文献4

  • 1GB11893-89.水质总磷的测定.钼酸铵分光光度法[S].2002.
  • 2DR/2500 Laboratory Spectrophotometer. Hach Company, 27 February 03, Allfights reserved. Printed in the U. S. A. Phosphorus,Total Method 8190 PhosVer3 with Acid Persulfate Digestion Method [ S ]. 2003.
  • 3国家环境保护总局.水和废水监测分析方法[M].北京:中国环境科学出版社,2002,12.144-153.
  • 4中国环境监测总站.环境水质监测质量保证手册[M].北京:化学工业出版社,1994.238-298.