2种标记方法在脑肿瘤CT/MRI图像配准中的比较 被引量:1

Comparison Between Two Marking Methods in CT/MRI Image Registration
下载PDF
导出
摘要 目的:比较放疗中CT/MRI图像配准时采用2种标记方法的差异。方法:获取9例脑肿瘤患者CT和MR图像时用不同种材料做匹配点(带绝缘皮电线和灌注釓喷酸葡胺注射液胶管);相同患者再用另一同种材料做匹配点(球形鱼肝油胶囊中心灌注碘海醇注射液),相同的CT和MRI扫描体位,分别获取另一组CT和MR图像,将所得图像用网络传输至治疗计划系统进行图像配准并评估结果。结果:CT和MRI定位扫描分别使用不同标记材料的一组最大配准一致性平均误差(CRE),范围为0.16~0.28cm;使用相同标记材料的一组配准一致性平均误差(CRE)范围为0.06~0.16cm。结论:使用能同时在CT和MRI上显影的相同标记材料做匹配点便于定位,且配准效果也更理想。 Objective To compare the results of CT/MRI registration by using two different marker materials. Methods CT and MR transversal images (using electric wire and thin tube filled with lohexol injection as matching points respectively) were obtained from nine cases with brain tumor. Another CT and MR transversal images were also obtained by using the cod-liver oil capsule filled with gadopentetate dimeglumine injection as matching points with the same scanning position. The images were transferred to the TPS for image registration and the effect was evaluated. Results The CT/MRI CRE range for using different kinds of marker material was 0.16-0.28cm, while was 0.06-0.16cm for using the same marker material. Conolution Comparing with the method using different kinds of material, the method using the same spherical material can get better results in CT/MRI registration.
出处 《医疗卫生装备》 CAS 2009年第8期71-72,共2页 Chinese Medical Equipment Journal
关键词 图像配准 CT MRI image registration CT MRI
  • 相关文献

参考文献5

  • 1Prott F J,Haverkamp U. Effect of distortion and asymmetry in MR images on radiotherapeutic treatment planning[J]. Int J Cancer, 2000,90 : 46-50.
  • 2付杰,胡超苏,胡伟刚,陆惠忠,何少琴.鼻咽癌CT/MRI配准方法的临床研究[J].中华放射肿瘤学杂志,2006,15(2):85-88. 被引量:7
  • 3Mah D,Steckner M,Hanlon A,et al. MRI simulation :effect of gradient distortions on three-dimensional prostate cancer plans[J]. Int J Radiat Oncol Biol Phys,2002,53:757-765.
  • 4徐利明,张照喜,刘永胜,王晓红,何正中.MRI直接用于颅内肿瘤治疗计划设计的可行性探讨[J].中华放射肿瘤学杂志,2005,14(4):327-329. 被引量:6
  • 5罗述谦 周果宏.医学图像处理与分析[M].北京:科学出版社,2003..

二级参考文献16

  • 1易俊林,高黎,黄晓东,李素艳,罗京伟,徐国镇.鼻咽癌放射治疗的失败模式[J].中华放射肿瘤学杂志,2004,13(3):145-148. 被引量:89
  • 2胡逸民 张红志 戴建荣 胡逸民 主编.治疗计划的设计与执行[A].胡逸民,主编.肿瘤放射物理学[C].北京:原子能出版社,1999.463-471.
  • 3Mah D,Steckner M,Hanlon A,et al.MRI simulation:effect of gradient distortions on three-dimensional prostate cancer plans.Int J Radiat Oncol Biol Phys,2002,53:757-765.
  • 4Lee N,Xia P,Fischbein NJ.Intensity-modulated radiation therapy for head-and-neck cancer:the UCSF experience focusing on target volume delineation.Int J Radiat Oncol Biol Phys,2003,57:49-60.
  • 5Poon PY,Tsang VH,Munk PL.Tumor extent and T stage of nasopharyngeal carcinoma:a comparison of magnetic resonance imaging and computed tomographic finding.Can Assoc Radiol J,2000,51:287-295.
  • 6Lau KY,Kan WK,Sze WM,etal.Accuracy of nasopharyngeal carcinoma staging by magnetic resonance imaging.Australas Radiol,2004,48:14-16.
  • 7Chung NN,Ting LL,Hsu WC,et al.Impact of magnetic resonance imaging versus CT on nasopharyngeal carcinoma:primary tumor target delineation for radiotherapy.Head Neck,2004,26:241-246.
  • 8Prott FJ,Haverkamp U,Eich H,et al.Effect of distortions and asymmetry in MR images on radiotherapeutic treatment planni.ng.Int J Cancer,2000,90:46-50.
  • 9Petra A,Vanden E,Evert-Jan DP,et al.Medical image matching-A review with classification.IEEE Eng Med Biol,1993,12:26-39.
  • 10Fitzpatrick JM,West JB,Maurer CR.Predicting error in rigid-body point-based registration.IEEE Trans Med Imaging,1998,17:694-702.

共引文献110

同被引文献7

引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部