摘要
采用多孔微球填充柱气相色谱外标顶空进样法,对部分一次使用性注射器及输液器上环氧乙烷残留量进行检测。结果,用环氧乙烷灭菌当天,产品上环氧乙烷残留量达2015~4396μg/g;放置14天,其残留量下降99%,放置30天者下降999%,但残留量仍有16~29μg/g。
The residual ethylene oxide on some disposable syringes and infusion sets was examined by gas chromatography The results indicated that the residual ethylene oxide on the products was 2015~4396 μg/g on the day of sterilization by ethylene oxide, decreased by 99% after storing for 14 days and by 999% after storing for 30 days, but the residual amount was still 16~29 μg/g
出处
《中国消毒学杂志》
CAS
1998年第3期176-179,共4页
Chinese Journal of Disinfection
关键词
注射器
输液器
一次性
环氯乙烷
残留量
disposable syringe disposable infusion set residual ethylene oxide gas chromatography