期刊文献+

皮层体感诱发电位在脊柱手术监护中的应用

下载PDF
导出
摘要 目的分析皮层体感诱发电位(Cortical Somatosensory Evoked Potential,CSEP)在脊柱手术监护中的应用价值。方法对48例脊柱手术患者术中行CSEP监护,并对结果进行比较和分析。结果48例中,10例无变化,其中1例术后出现一侧下肢轻瘫,24小时后恢复如常。38例术中CSEP发生改变,其中30例CSEP改变但波幅下降<50%,潜伏期延长<10%;3例波幅下降>50%和/或潜伏期延长>10%,其中1例唤醒试验阳性;5例明显改善。结论术中CSEP监护可对医源性脊髓神经损伤及时发出预警,可有效降低手术后肢体瘫痪等神经功能损害的发生率。
机构地区 丽水市中心医院
出处 《浙江实用医学》 2009年第4期314-315,333,共3页 Zhejiang Practical Medicine
  • 相关文献

参考文献6

  • 1Nash C L.Jr,Loming R A, Schatzinger L A,et al. Spinalcord monitoring during operation treatment of the spine. Clin Orthop, 1977, 126: 100.
  • 2Luk K D, Hu Y, Wong Y W, et al. Variability of sornatosensory - evoked potentials in differentstages of scoliosis surgery. Spine, 1999, 24:1799.
  • 3Salzman S K, Dabney K W, Mendez A A,et al.The somatosensory evoked potential predicts neurologic deficits and erotonergic pathochemistry after spinal distraction injury in experimental scoliosis. J Neurotrauma, 1988,5(3) : 173.
  • 4王德春,齐宗华,胡有谷.脊髓监护在脊柱手术中的应用[J].中国脊柱脊髓杂志,2004,14(9):569-571. 被引量:9
  • 5胡勇,胡从云,陆瓞骥.脊柱侧凸矫形术中脊髓监护基准的选择[J].中华骨科杂志,2000,20(9):555-558. 被引量:37
  • 6Nuwer M R. Use of sematosensory evoked potentials for intraoperafive monitoring of cerebral and spinal cord function. Neurol Clin, 1988,6 (4) :881.

二级参考文献23

  • 1Legatt AD.Current practice of motor evoked potential monitoring:results of a survey[J].J Clin Neurophysiol,2002,19(5):454-460.
  • 2MacDonald DB. Safety of intraoperative transcranial electrical stimulation motor evoked potential monitoring[J].J Clin Neurophysiol, 2002,19 ( 5 ) :416-429.
  • 3Weber M, Eisen AA. Magnetic stimulation of the central and peripheral nervous systems[J].Muscle Nerve,2002,25(2):160-175.
  • 4York DH. A critical evaluation of the 50% criterion in SEP monitoring[C]. Paper Presented at the Second Annual Meeting of the American Society for Neurophysiological Monitoring.Pittsburgh: PA, 1991.
  • 5Kombos T, Suess O, Da Silva C,et al.Impact of somatosensory evoked potential monitoring on cervical surgery[J].J Clin Neurophysiol, 2003,20 ( 2 ): 122-128.
  • 6Hu Y, Luk KD, Lu WW,et al. For intraoperative spinal cord monitoring [J].J Neurol Neurosurg Psychiatry,2003,74 ( 1 ):82-87.
  • 7Pelosi L, Lamb J,Grevitt M,et al. Combined monitoring of motor and somatosensory evoked potentials in orthopaedic spinal surgery[J].Clin Neurophysiol,2002,113 (7):1082-1091.
  • 8Jou IM, Chern TC, Chen TY, et al. Effects of desflurane on spinal somatosensory-evoked potentials and conductive spinal cord evoked potential[J].Spine,2003,28(16):1845-1850.
  • 9Jones SJ,Buonamassa S,Crockard HA.Two cases of quadriparesis following anterior cervical discectomy,with normal perioperative somatosensory evoked potentials [J].J Neurol Neurosurg Psychiatry, 2003,74 (2):273-276.
  • 10Jones SJ,Harrison R,Koh KF,et al. Motor evoked potential monitoring during spinal surgery:response of distal limb muscles to transcranial cortical stimulation with puls trains[J].Electroenceph Clin Neurophysical,1996,100(5):375-383.

共引文献43