摘要 适用于淡水的BOD测定仪已经成型,海水盐度高,BOD浓度低,测定仪样品传输及测量的精度和微生物传感器菌种的耐盐性,给构建海水BOD自动测定仪带来困难。以溶解氧电极微生物传感器法为基础,依据海水BOD测定的技术特点,分析样品传输系统、恒温系统和信号采集与处理系统的技术要求,并筛选耗氧耐盐菌种作为微生物传感器的菌株,构建了海水BOD测定仪。用该测定仪测定海水标样,测定结果与标准稀释法测定的BOD5具有良好的相关性和准确度。 The B0D instrument for fresh water has been forming, High salinity of sea water,the low concentration of BOD, the accuracy of samples transmission and measurement, the salt-tolerant of microbial sensors bacteria, is the difficult to build auto-testing instrument for seawater BOD. This research is based on dissolved oxygen electrode microbial sensor, base on the technical characteristics of the determination of seawater's BOD, analysis samples transmission system, temperature system, signal acquisition and processing system, screening of oxygen salt-tolerant bacteria as microbial sensors bacteria, to build a auto-determining instrument for seawater's BOD. This instrument determines seawater's BOD, whose results were agreed well with those by standard method (BOD5).
出处 《中国环境监测》 CAS CSCD 北大核心 2009年第4期64-66,共3页 Environmental Monitoring in China
基金 国家"863"计划海洋技术领域项目(2007AA092201)
关键词 海水 BOD 微生物传感器 测定仪 BOD Seawater Microbial sensor Auto-determining instrument
  • 相关文献

参考文献7

二级参考文献16

  • 1许春向,马文科,杜晓燕.化学量传感器 第五讲 生物传感器 七 生物传感器的应用[J].传感器技术,1993(6):45-51. 被引量:4
  • 2胡国强 杨惠玲.生化需氧量测定技术的进展[J].环境科学丛刊,1985,6(12):8-13.
  • 3[1]美国公共卫生协会水污染控制联合会等编,宋仁元等译.水和废水标准检验法(第15版)[M].重庆:中国建筑工业出版社,1998.444-449.
  • 4[2]奚旦立等.环境监测(修订版)[M].北京:高等教育出版社,1994.83-84.
  • 5.ISO/TC147.水质5日生化需氧量(BOD5)的测定,稀释与接种法[Z].,..
  • 6奚旦立 等.环境监测(修订版)[M].北京:高等教育出版社,1994..
  • 7李友荣,生物工程学报,1989年,5卷,235页
  • 8匿名著者
  • 9Li Y,Appl Biochem Biotechnol,1991年,28/29卷,855页
  • 10Zang X,环境科学学报,1986年,6卷,184页

共引文献61

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部