^(32)P持续低剂量率辐射敷贴治疗的胶片剂量验证 被引量:3

The quantitative radiation dose by RC film for the continued low-dose-rate radiation of ^(32)P
下载PDF
导出
摘要 目的:对32P的敷贴治疗进行胶片剂量验证。方法:直线加速器建立25-575cGy的胶片辐射剂量曲线。放射性核素32P溶液4.14MBq吸附在1.1cm×1.2cm的长方形滤纸上制成放射性敷贴器。将20张1.5cm×1.5cm的辐射直接显色(RC)胶片(胶片厚0.225mm)重叠放置(总厚度4.5mm),把放射性敷贴器放置在最顶端,辐射0.5小时。通过辐射剂量曲线计算每张胶片辐射剂量,根据胶片厚度推算距离敷贴器不同距离处的辐射剂量,并绘制相应的剂量-距离曲线。结果:距离敷贴器0、0.225、0.45、0.675、0.9、1.125、1.35、1.8、2.025、2.25、2.475mm处的0.5小时辐射测量吸收剂量为265、175、120、98、68、55、39、25、19、16、12、5cGy,拟合曲线为Y=257.28e-1.4521X。结论:放射性核素32P敷贴器的辐射剂量随距离增加呈指数方式衰减,敷贴器表面和2.5mm处的辐射剂量相差约53倍。 Objective:To study the distribution of radiation dose in space of ^32P by EBT films. Methods: The curve of radiation dose and grey scale of film from 25 to 575cGy was built by linear accelerator. The radionuclide of ^32P was dripped on 1. 1cm × 1.2cm filter paper. The twenty 1.5cm × 1.5cm RC films ( thick film 0. 225mm ) were overlap together ( total thickness 4.5mm) and the filter paper of ^32P were placed on the top. There were 0.5 hours irradiation and the twenty films radiation dose were measured by standard curve. According to the thickness of film and radiation dose, the curve of radiation dose and distance was distributed. Results: The absorbed dose of the films at 0, 0. 225 ,0. 45 ,0. 675,0.9,1. 125,1.35,1.8,2. 025,2.25,2. 475mm place were 265,175,120,98,68,55,39,25,19, 16,12,5 cGy. The fitting curve was Y = 257.28e^- 1 4521X. Conclusion: The radiation dose of radionuclide 32p were decayed in distance by index mode and there was 53 times difference between the surface and 2.5mm distance of the ^ 32P.
出处 《现代肿瘤医学》 CAS 2009年第9期1772-1774,共3页 Journal of Modern Oncology
关键词 32P RC胶片 剂量验证 ^ 32P RC film dose verification
  • 相关文献

参考文献5

  • 1施常备,李竟成,许建林.^(32)P治疗婴幼儿皮肤血管瘤61例[J].陕西医学杂志,2005,34(2):233-234. 被引量:8
  • 2王世真,周前.核医学[M].上海:上海科学技术出版社,1992:4-10.
  • 3Ehsan Samei, Kimberlee J Kearfortt,Timothy J Gllespie,et al. An atlas of selected bete - ray spectra and depth - dose distributions in lithium fluoride and soft tissue generated by a fast monter carlo - based sampling method[ J]. Radiat Phys Chem, 1996,6 (48) :719 - 725.
  • 4Wesley E Bolch, Lionel G Bouchet, James S Robertson, et al. The dosimetry of nonuniform activity distributions - radionuelide S values at the voxel level [J]. J Nuel Med, 1999,40:11s- 36s.
  • 5周颖娟(综述),黄劭敏(综述),邓小武(综述).辐射显色剂量胶片的特性及其在放射医学的应用[J].国际肿瘤学杂志,2007,34(1):28-31. 被引量:8

二级参考文献19

  • 1赵惠杨.校医学[M].上海:上海科学技术出版社,1981.412.
  • 2Chiu-Tsao ST,Ho Y,Shankar R,et al.Energy dependence of response of new high sensitivity radiochromic films for megavoltage and kilovoltage radiation energies.Med Phys,2005,32(11):3350-3354.
  • 3Cheung T,Butson MJ,Yu PK.Post-irradiation colouration of Gafchromic EBT radiochromic film.Phys Med Biol,2005,50 (20):N281-285.
  • 4Sharma SD,Bianchi C,Conte L,et al.Radiochromic film measurement of anisotropy function for high-dose-rate Ir-192 brachytherapy source.Phys Med Biol,2004,49(17):4065-4072.
  • 5Piermattei A,Fidanzio A,Azario L,et al.A standard dosimetry procedure for 192Ir sources used for endovascular brachytherapy.Phys Med Biol,2002,47 (23):4205-4221.
  • 6Kirisitsa C,Stembergera A,PokrajacaB,et al.Clinical quality assurance for endovascular brachytherapy devices.Radiother Oncol,2004,71(1):91-98.
  • 7Butson MJ,Cheung T,Yu PK,et al.Assessment of large single-fraction,low-energy X-ray dose with radiochromic film.Int J Radiat Oncol Biol Phys,2000,46(4):1071-1075.
  • 8Gamble LM,Farrell TJ,Jones GW,et al.Composite depth dose measurement for total skin electron (TSE) treatments using radiochromic film.Phys Med Biol,2003,48(7):891-898.
  • 9Kubo HD,Araki F.Dosimetry and mechanical accuracy of the first rotating gamma system installed in North America.Med Phys,2002,29(11):2497-2505.
  • 10Daftari I,Castenadas C,Petti PL,et al.An application of GafChromic MD-55 film for 67.5 MeV clinical proton beam dosimetry.Phys Med Biol,1999,44(11):2735-2745.

共引文献14

同被引文献21

引证文献3

二级引证文献3

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部