影像引导放疗中千伏特锥形束CT患者剂量的测量 被引量:1

Patient dose evaluation for kilovoltage cone beam CT in image-guided radiotherapy
原文传递
导出
摘要 目的测量医用直线加速器机载千伏特锥形束CT扫描过程中患者的剂量。方法用PTW TM30009电离室分别在空气中、T40017头部和T40016躯干模体中,改变XⅥ锥形束CT的管电压、管电流、准直器以及机架旋转范围等参数测量空气CT剂量指数和加权CT剂量指数,计算相应的剂量长度乘积,并与ACQSim模拟定位CT的结果进行比较。结果千伏特锥形束CT在空气中和模体中的剂量指数随管电压二次方变化,随管电流线性变化。加权剂量指数和剂量长度乘积与准直器以及机架旋转范围有关。临床常用条件下千伏特锥形束CT归一化的剂量指数比模拟定位CT小。结论CT剂量指数能反映成像过程中患者接受的剂量,测量方法简单、重复性好,应定期进行测量。锥形束CT扫描时应该合理选择成像参数,尽量减小扫描范围,最大限度减少患者接受剂量。 Objective To evaluate patient dose from linac-integrated kilovolt cone beam CT (CBCT) imaging in image-guided radiotherapy. Methods Air CT dose index (CTDIair) and weighted CT dose index (CTDIw ) were measured with PTW TM30009 CT ion chamber in the air, in T40017 head phantom and T40016 body phantom, respectively. During the measurement, different combinations of tube voltage, current, exposure time per frame, collimator, and gantry rotation range of XⅥ CBCT were used. Dose length products (DLP) were subsequently calculated. Normalized CTDI was then compared to ACQSim simulator CT. Results Both CTDIair and CTDIw depended quadratically on the voltage, while linearly on mAs. CTDIw and DLP had close relation with the collimator and the gantry rotation range. Normalized CTDIw for X VI CBCT under routine clinical condition was lower than that of ACQSim simulator CT . Conclusions CTDIw, being convenient and reliable for measurement, can be used to assess the patient dose during CBCT imaging. Regular measurement should be performed in QA & QC program. Optimal image parameters should be chosen to reduce the scanning range and patient dose.
出处 《中华放射肿瘤学杂志》 CSCD 北大核心 2009年第5期405-408,共4页 Chinese Journal of Radiation Oncology
关键词 放射疗法 影像引导 断层摄影术 X线计算机 锥形束 断层摄影术 X线计算 剂量指数 剂量长度乘积 Radiotherapy, image-guided Tomograph, X-ray computed, cone beam Tomograph, X-ray computed, dose index Dose length product
  • 相关文献

参考文献8

  • 1Oelfke U, Tucking T, Nill S, et al. Linac-integrated kV-cone beam CT: technical features and first applications. Med Dosim, 2006, 31:52-70.
  • 2Murphy M J, Baiter J, Balter S, et al. The management of imaging dose during image-guided radiotherapy:report of the AAPM Task Group 75. Med Phys,2007,34:4041-4063.
  • 3Islam MK, Purdie TG, Norrlinger, et al. Patient dose from kilovoltage cone beam computed tomography imaging in radiation therapy. Med Phys,2006,33:1573-1582.
  • 4Lehmann J, Perks J, Semon S, et al. Commissioning experience with cone-beam computed tomography for image-guided radiation therapy. J Appl Clin Med Phys,2007 ,8 :21-36.
  • 5Tsapaki V, Aldrich JE, Sharma R, et al. Dose reduction in CT while maintaining diagnostic confidence : diagnostic reference levels at routine head, chest, and abdominal CT-IAEA coordinated research project. Radiology,2006, 240:828-834.
  • 6International Commission on Radiation Units and Measurement. Patient dosimetry for X-rays used in medical imaging. ICRU Report 74. Genevese:ICRU,2005.
  • 7Stutzel J, Oelfke U, Nill S. A quantitative image quality comparison of four different image guided radiotherapy devices. Radiother Oncol,2008,86:20-24.
  • 8Amer A, Marchant T, Sykes J, et al. Imaging doses from the Elekta Synergy X-ray cone beam CT. Br J Radiol.2007 ,80 :476-482.

同被引文献4

引证文献1

二级引证文献6

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部