MatriXX系统在放疗质控中的应用研究 被引量:11

Research of the Application of MatriXX System in Radiation Therapy Quality Control
原文传递
导出
摘要 目的探讨二维空气电离室矩阵MatriXX系统进行放疗质控监测的准确性、可靠性和方便性。方法用MatriXX系统在固体水模块中测量医用加速器X射线射野的百分深度剂量(PDD)、组织最大比(TMR)、离轴比、半影和射野宽度,并与RFA300三维水箱扫描结果进行对比分析。MatriXX与自制剂量测量体模相结合验证Pinnac3V8.0m三维放疗计划系统设计的静态调强和动态调强放疗计划。结果MatriXX在固体水中所测6MV X射线的PDD和TMR曲线与RFA300水箱扫描曲线基本平行,15MV X射线的PDD曲线与三维水箱扫描结果几乎重合。用MatriXX所测射野半影比三维水箱扫描值大两倍左右,当射野小于3cm×3cm时MatriXX不能自动计算出半影值。MatriXX可在最短采样时间20mS内同时获得射野离轴比、对称性、平坦度、野宽等信息,并能快速准确地完成调强放疗计划二维剂量分布的定量和定性验证。结论MatriXX在放疗质控测量时具有准确、简便、高效的优势,是较为理想的放疗质控工具。 Objective To investigate the accuracy, reliability and convenience of 2 - D ionization chamber array MatriXX system in daily quality control of radiation therapy. Methods The percent depth dose (PDD) , tissue maximum ratio (TMR) , off-axis ratio profile, penumbra and field width of Varian 23EX linac were measured by MatriXX system combined with solid water phantom and compared with the data acquired by RFA300 three - dimensional water phantom. The dose distri- butions of static and dynamic intensity modulated radiation therapy (IMRT) plans created by Pinnac3 V8.0m three - dimen- sional treatment planning system (3D -TPS) were validated by MatriXX combined with the dose phantom made in home. Results The PDD and TMR of 6MV X - ray measured by MatriXX system were parallel to the curves acquired by RFA300, and the PDD curve of 15MV X - ray measured by MatriXX was nearly identic to that measured by RFA300. The values of penumbra measured by MatriXX were two or three times larger than those measured by RFA300, and when the field size was smaller than 3cm×3cm the penumbra of it could not be automatically calcuiatod out by the MatriXX system. The characteristic information of field, such as symmetry and flatness and field width, can be acquired at once in a single sampling, which can be completed in less than 20mS. The qualitative and quantificational verification of 2 - D close distribution of IMRT plans can be quickly implemented by using MatriXX system. Conclusions The MatriXX system is very accurate and convenient and effective in quality control of clinical radiotherapy, and it' s a perfect tool for the quality control procedure of radiation therapy.
出处 《中国辐射卫生》 2009年第3期272-274,共3页 Chinese Journal of Radiological Health
基金 广东省科技计划项目(2007B031509009) 广州医学院自然科学科研项目(0702009)
关键词 剂量验证 质量控制 二维空气电离室矩阵 MatriXX系统 Dose Verification Quality Control Two - Dimensional Ionization Chamber Array MatriXX System
  • 相关文献

参考文献7

  • 1张书旭,周凌宏,陈光杰,林生趣,沈国辉,王余峰.二维空气电离室矩阵MatriXX系统的应用研究[J].中华放射医学与防护杂志,2009,29(1):93-96. 被引量:17
  • 2张书旭,徐海荣,陈光杰,杨克柽,林生趣,余辉.呼吸运动对靶区三维重建的影响[J].中国医学物理学杂志,2006,23(6):400-404. 被引量:14
  • 3Andenna C, Benassi M, Caccia B, et al. Comparison of dose distributions in IMRT planning using the gamma function [ J ]. J Exp Clin Cancer Res,2006,25 (2) :229 - 234.
  • 4Amerio S, Boriano A, Bourhaleb F, et al. Dosimetric characterization of a large area pixel - segmented ionization chamber [J]. MedPhys,2004,31(2):414-420.
  • 5Herzen J, Todorovic M, Cremers F, et al. Dosimetric evaluation of a 2D pixel ionization chamber for implementation in clinical routine[J]. Phys Med Biol,2007,52(4) : 1 197 - 1 208.
  • 6Jursinic PA, Nelms BE. A 2 - D diode array and analysis software for verification of intensity modulated radiation therapy delivery [ J ]. Med Phys, 2003,30 ( 5 ) : 870 - 879.
  • 7Wiezorek T, Banz N, Schwedas M. Dosimetric quality assurance for intensity - modulated radiotherapy feasibility study for a filmless approach [ J ]. Strahlenther Onkol, 2005, 181 (7) :468 -474.

二级参考文献10

  • 1于甬华,吴玉芬,郭守芳,王仁本,李文武,李宝生.呼吸运动对周围型肺癌位置的影响及其数学模型的建立[J].中华放射肿瘤学杂志,2004,13(2):83-84. 被引量:20
  • 2Amerio S, Boriano A, Bourhaleb F, et al. Dosimetric 'characterization of a large area pixel-segmented ionization chamber. Med Phys, 2004, 31(2): 414-420.
  • 3Herzen J, Todorovic M, Cremers F, et al. Dosimetfic evaluation of a 2D pixel ionization chamber for implementation in clinical routine. Phys Med Bid, 2007, 52(4): 1197-1208.
  • 4Gossman MS, Robertson MA, Lawson RC. Correlation between detector array measurements and a computer algorithm for enhanced dynamic wedge profiles. Med Dosim, 2007, 32(3): 211-215.
  • 5Buonamici FB, Compagnucci A, Marrazzo L, et al. An intercomparison between film dosimetry and diode matrix for IMRT quality assurance. Med Phys, 2007, 34(4) : 1372-1379.
  • 6Andenna C, Benassi M, Caccia B, et al. Comparison of dose distributions in IMRT planning using the gamma function. J Exp Clin Cancer Res, 2006, 25(2): 229-234.
  • 7Erridge SC,Sepenwoolde Y,Muller SH,et al.Portal imaging to assess set-up errors,tumor motiom and tumor shrinkage during conformal radiotherapy of non-small cell lung cancer[J].Radither Oncol,2003,66(1):75-85.
  • 8黄伟,李宝生,于金明.精确放疗中呼吸运动对肺部肿瘤靶区的影响与控制[J].国际肿瘤学杂志,2005,32(1):53-56. 被引量:25
  • 9李宝生,於文雪,尹勇,李松毅,于金明.应用模拟定位机对肺癌肿块运动的研究[J].东南大学学报(自然科学版),2002,32(4):590-593. 被引量:19
  • 10于甬华,罗立民,李宝生,于金明,梁超前,赵月环.呼吸运动对肝部占位性病灶位置的影响及数学模型的建立[J].中华放射肿瘤学杂志,2002,11(4):245-247. 被引量:25

共引文献28

同被引文献65

引证文献11

二级引证文献42

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部