NEMA NU 2-2001标准在国内PET/CT性能测试的整体应用 被引量:7

NEMA NU 2-2001 Standards Used for PET/CT Performance Tests in China
下载PDF
导出
摘要 目的应用NEMA NU 2-2001标准测试国内部分PET/CT的性能指标。方法按照NEMA NU 2-2001标准的测试模型和方法,测试PET空间分辨率、散射系数、计数丢失和随机符合性能、灵敏度以及成像质量、衰减和散射校正精度性能指标,探讨PET/CT融合精度测试方法,并对LSO晶体PET/CT的本底计数率特性作初步测试。结果各测试指标均在厂家标称值范围内。结论采用标准的NEMA测试模型是PET性能测试的重要环节,其次是要严格控制测试模型的放射性活度,NEMA NU 2-2001标准中使用的模型或方法对于大多数使用者是可行的,可用于PET/CT安装前后的验收测试,部分测试可用于质量控制。 Objective To test the performances of PET/CT scanner in China with NEMA NU 2-2001 Methods Using the test phantoms and methods according to NEMA NU 2-2001,the spatial resolution; scatter fraction, count losses, randoms measurement, sensitivity; image quality, accuracy of attenuation and scatter corrections were measured. The measures of image fusion precision was discussed and premilary measurement to background count of LSO crystal was done. Results The results of measurment were all in the range of factory standard. Conclusion Using the standard NEMA phantoms and controlling the lowest activity level of test source (the count rate of test source) are critical to test the PET/CT performance..The NEMA NU 2-2001 is available to acceptance test of PET/CT scanner in pre- and post-installation periods, and a part of test can be used in the quality control.
出处 《中国医疗设备》 2009年第8期28-32,共5页 China Medical Devices
关键词 NEMA NU 2-2001标准 PET/CT 性能测试 NEMA NU 2-2001 standards PET/ CT performance tests
  • 相关文献

参考文献10

二级参考文献42

  • 1冯彦林,贺小红,黄克敏,余丰文,刘德军,袁建伟,袁白虹,苏少弟.Gemini PET/CT系统性能模型测试[J].中华核医学杂志,2005,25(1):54-56. 被引量:9
  • 2李小华,贺小红,李凡勇,陈萍,陶军,刘炜.Discovery ST PET性能测试与NEMA标准的应用[J].中华核医学杂志,2006,26(2):104-106. 被引量:14
  • 3宋金龙,邵文博,丛宁.PET-CT同机融合显像在肝癌TACE治疗后的应用价值[J].实用放射学杂志,2007,23(1):109-111. 被引量:4
  • 4Tsuyoshi K, Eiji O, Tomio I. Deep-inspiration breath-hold PET/CT of lung cancer: maximum standardized uptake value analysis of 108 patients [J]. J Nuel Med, 2008,49(8):1223-1231.
  • 5Karger CP, Hipp P, Henze M, et al. Stereotaetie imaging for radiotherapy: accuracy of CT, MRI, PET and SPECT [J]. Phys Med Biol, 2003,48(2):211-221.
  • 6National Electrical Manufactures Association, NEMA NU2-2001. Performance measurements of positron emission tomographs [S ]. NEMA Standards Publication, 2001.
  • 7Watson CC, Casey ME, Eriksson L, et al. NEMA NU 2 performance tests for scanners with intrinsic radioactivity [J ]. J Nucl Med, 2004, 45(5): 822-6.
  • 8Erdi YE, Nehmeh SA, Mulnix T, et al. PET performance measurements for an LSO-based combined PET/CT scanner using the national electrical manufacturers association NU 2-2001 standard[J]. J Nucl Med, 2004, 45(5): 813-21.
  • 9Kadrmas DJ, Christian PE. Comparative evaluation of lesion detectability for 6 PET imaging platforms using a highly reproducible whole-body phantom with (22)Na lesions and localization ROC analysis[J]. J Nucl Med, 2002, 43( 11 ): 1545-1554.
  • 10Bettinardi V, Pagani E, Gilardi MC. Implementation and evaluation of a 3D one-step late reconstruction algorithm for 3D positron emission tomography brain studies using median root prior[J]. Eur J Nucl Med Mol Imag, 2002, 29 ( 1 ): 7-18.

共引文献44

同被引文献41

引证文献7

二级引证文献25

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部