Quality Assurance for RuiDi RD/DB Carbon Fiber Immobilization Body-Frame
下载PDF
导出
摘要 目的:对新购进瑞迪RD/DB碳纤维定位体架进行全面检验,建立相应的质量保证体系。方法:(1)肉眼外观检查体架一般情况;使用钢卷尺测量外表坐标刻度的准确性。(2)在模拟机下使用激光灯检测左右Y轴,Z轴外表刻度的一致性。(3)使用直线加速器6MVX-线,固体水模及0.66cc电离室测量体架穿射因子随测量深度及机架角度变化情况。(4)使用模体模仿实际情况检测内置坐标同外表刻度的一致性。结果:卷尺测量右侧Y轴,右侧Z轴及X轴均无差异;Y轴外表坐标刻度400~100范围左侧较右侧数值最大范围小1mm,800~400间无差异;Z轴左右对应无差异;X轴100外表刻度对应内置坐标之定位基准线;穿射因子随测量深度和机架角增加而增加,机架角15°以前变化较缓,后变化加快;1mm,3mm,5mm,7mm层厚CT扫描传入TPS重建并计算得等中心坐标同预设坐标均有差异,最大值<1.8mm。结论:实际检测结果符合临床治疗要求,但仍须对新购体架作全面验证,实际计算出等中心坐标需经模拟机或EPID验证后方能作为治疗坐标使用。 Objective: To check the new-purchased RuiDi RD/DB carbon fiber immobilization body-frame to set up the related quality assurance system. Methods: (1)The RuiDi body-flame's appearance was evaluated objectively and its outside coordinates was verifide measured to check their accuracy by using a steel ruler. (2)The agreement of outside coordinates for Y-axis and Z-axis were checked by using simulator laser. (3)The body-flame transmission factor was measured in solid water phantom using 0.6 cc ion chamber for 6 MV X-ray .Absorbed dose fluctuation with different measured depth and gantry angle was also measured. (4)The agreement of outside and inner coordinates system was also verified by simulating actual treatment process in which the solid water phantom was also used .Results: The scale of right outside Y, Z, X coordinates are accurate. The left outside Y-axis scale is 1 mm shorter than the right one between 400-100 while there is no difference between them for the range 400-800.The transmission factor increases with the measured depth and gantry angle's increase.It changes slowly with the gantry angle less than 15 degree.The CT studies of 1 mm,3 mm,5 mm,7 mm slices were transfered to the TPS to be reconstructed. There are differences between all the measured ISO-coordinates and the preset coordinates, but the maximum error is less than 1.8 mm. Conclusions: The checked results for RuiDi RD/DB carbon fiber immobilization body-frame meet our clinical radiotherapy requirement after recalibrated. However, Every newly purchased immobilization body-frame should be checked comprehensively .The actual measured treatment ISO-coordinates can not be used until they're verified by using simulator or EPID.
出处 《中国医学物理学杂志》 CSCD 2009年第5期1384-1388,共5页 Chinese Journal of Medical Physics
关键词 体架 质量保证 body-frame quality assurance
  • 相关文献

参考文献4

  • 1Tsang Cheung, Martin J. Butson, Peter K.N.Yu. Evaluation of buildup dose from 6 MV X-ray under pelvic and abdominal patient immobilization devices [J]. Radiation measurement, 2002,35:235-238.
  • 2James M. Galvin,D.Sc., Greg Bednarz,Ph.D. Quality assurance procedures for stereotactic body radiation therapy [J]. Int J Radiat Oncol Biol Phys, 2008,71:122-125.
  • 3李兆斌,徐利明,王晓红,李承军.ELEKTA立体定向体架的质量保证和质量控制[J].肿瘤防治研究,2006,33(11):843-844. 被引量:3
  • 4沙翔燕,王运来,廖雄飞,解传滨,戴相昆,马林.全碳素纤维治疗床对吸收剂量的影响[J].中华放射肿瘤学杂志,2008,17(3):223-225. 被引量:11

二级参考文献9

  • 1Myint WK, Niedbala M, Wilkins D. Investigating treatment dose error due to beam attenuation by a carbon fiber tabletop. J Appl Clin Med Phys ,2006,7:21-27.
  • 2Meara SJP, Langmack KA. An investigation into the use of carbon fiber for megavoltage radiotherapy applications. Phys Med Biol, 1998,43:359-1366.
  • 3McCormack S, Diffey J, Morgan A. The effect of gantry angle on megavoltage photon beam attenuation by a carbon fiber couch insert. Med Phys ,2005 ,32 :483-487.
  • 4任妍,陈林.体部固定框架在三维立体定向适形放疗中的应用[C].首届全国放射治疗技师大会论文集-质量控制部分[M].上海:复旦大学附属肿瘤医院,2005.366-367.
  • 5S Yoon,E Choi,S Lee,et al.Stereotactic Body Frame Based Fractionated Radiosurgery for Primary or Metastatic Thoracic Tumors[J].Int J Radiation Oncology Biol Phys,2004,60(Supplement):s282.
  • 6Ronald C,Mcgarry,Lech Papiez,et al.Stereotactic Body Radiation therapy of early stage non-small-cell lung carcinoma:phase I study[J].Int J Radiation Oncology Biol Phys,2005,63 (4):1010-1015.
  • 7JOrn Wulf,Ulrich H(a)dinger,Ulrich Oppitz,et al.Stereotactic radiotherapy of extracranial targets:CT-simulation and accuracy of treatment in the stereotactic body frame[J].Radiotherapy and Oncology,2000,57(2),225-236.
  • 8杨树松,王凡,孙维凯,刘建海,刘启平,李文.治疗床面下垂导致的系统摆位误差分析及对策[J].中华放射肿瘤学杂志,2001,10(1):55-57. 被引量:45
  • 9单国平,严英师,徐明清.固定面罩对射线的剂量学影响[J].中华放射医学与防护杂志,2003,23(2):112-113. 被引量:11

共引文献12

同被引文献1

引证文献1

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部