期刊文献+

呼吸机吸气压力示值误差的检测方法研究和评定 被引量:4

Design and evaluation of measuring method for error of ventilator inspiratory pressure indication
下载PDF
导出
摘要 目的:设计呼吸机吸气压力示值误差的检测方法。方法:首先分析呼吸机产品和临床救治中需要的吸气压力范围和最大允许误差,再以法制计量和计量学的要求来分析、确定呼吸机吸气压力检测仪器的测量范围和相应的最大允许误差。结果:通过测量实验和测量结果不确定度评定来分析、验证和确定呼吸机吸气压力示值误差的测量方法。结论:该测量方法符合计量学量值传递的要求,适合对呼吸机吸气压力设定值和监测值进行检测和质量控制的需要。 Objective: To design and evaluation ventilator inspiratory pressure indication error measurement method. Methods: First analyze the inspiratory pressure range and maximum allowable error in clinical treatment and the ventilator-products features. Second determine measuring range and maximum allowable error of the ventilator inspiratory pressure measuring instrument according to metrology regulations and theory. Results: Then analyze, verify and identify measuring method of ventilator inspiratory pressure indication error by measuring experiment and measurement uncertainty evaluation. Conclusion: Finally make sure the measurement method has sufficient accuracy, is in line with the metrology quantity value transfer requirements and match the needs of ventilator inspiratory pressure indication error measurement and monitor in clinical treatment.
作者 陈建功
出处 《中国医学装备》 2009年第10期9-13,共5页 China Medical Equipment
关键词 呼吸机 吸气压力示值误差 测量方法设计 测量不确定度 Ventilator Inspiratory pressure indication error Design of measuring method Uncertainty of measurement
  • 相关文献

参考文献6

  • 1GB9706.28-2006,医用电气设备(第2部分):呼吸机安全专用要求治疗呼吸机[S].
  • 2GB9706.29-2006,医用电气设备(第2部分):麻醉系统的安全和基本性能专用要求[S].
  • 3刘大为.危重病医学[M].北京:中国协和医科大学出版社,2000.211.
  • 4肖胜春,卢兴平,曹德森,刘延武.呼吸机通气质量检测及质量评估方法[J].中国医疗设备,2008,23(1):40-43. 被引量:34
  • 5周丹 曹德森 刘尚军 等.现代呼吸机应用的质量管理.临床工程理论与实践,2004,.
  • 6JJF1059-1999.测量不确定度评定与表示[S].[S].,..

二级参考文献5

共引文献893

同被引文献31

引证文献4

二级引证文献34

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部