摘要
金宝AK200自检能正常通过,但干粉桶进水不正常,然后电导报警。这是血透机使用过程经常会碰到的故障。文章根据图解详细分析了血透机的工作原理,根据故障现象准确查出故障原因所在,对同类设备的维修具有一定的指导意义。
The GAMBRO AK200 warning about the conductivity after selftest because water can't prime into the BI-CART. We'll find this wrong component frequently. This article analyzed the operation principle of this equipment in detail based on its electric circuit, and located the wrong component rapidly according its trouble phenomenon. We hope this could be a help in the maintenance of the same kind of equipments.
出处
《中国医学装备》
2009年第10期50-51,共2页
China Medical Equipment
关键词
电导
干粉桶
O形圈
滤网
阀
Conductivity
Bi-Cart
O-ring
Filter
Valve