期刊文献+

形变医用镍钛合金表面TiN涂层耐蚀性的研究

Study on Corrosion Resistance of TiN Coated Biomedical Nitinol under Deformation
下载PDF
导出
摘要 在37℃的人造血浆中,研究了有无TiN涂覆层的镍钛合金的耐蚀性。电化学实验显示TiN涂层能使其相对饱和甘汞电极的电位从0.95 V增至1.5 V,其腐蚀电流密度从0.20 mA/cm2降至0.10 mA/cm2。溶解的钛离子和镍离子的化合价为正二价。变形量小于6%时,镍钛基TiN膜层的耐蚀性并未受到严重影响。由于变形产生的TiN膜层裂纹而导致了轻微的腐蚀,但其耐蚀性仍好于没有TiN膜层的镍钛合金,因为在裂纹处未观察到电偶腐蚀。 The corrosion resistance of nitinol with and without a titanium nitride coating is electrochemically investigated at 37 ℃ in artificial plasma. The electrochemical experiments show that the titanium nitride (TiN) coating can increase the stable potential from 0.95 V to 1.5 V versus a saturated calomel electrode and reduce the corrosion current density from 0.20 mA/cm^2 to 0. 10 mA/ cm^2. The valence of the dissolved titanium and nickel ions is plus two. At deformation less than 6%, no significant negative effects on the corrosion resistance of the TiN coated nitinol is observed. More deformation causes cracks in the TiN coatings and consequently, the corrosion resistance degrades slightly, but it is still better than that of the uncoated nitinol since no galvanic corrosion is observed from the broken areas.
出处 《江西科学》 2009年第5期670-674,共5页 Jiangxi Science
关键词 氮化钛 镍钛合金 耐蚀性 Titanium nitride, Nitinol, Corrosion resistance
  • 相关文献

参考文献2

二级参考文献13

  • 1Castelman LS and Motzkin SM.Biocompatibility of clinical implants materials.ed,D F Williams (CRC Press,Boca Raton,1981) 1:129-154.
  • 2Wever DJ and Veldhuizen AG,et al.Electrochemical and surface characterization of anickel-titanium alloy.Biomaterials,1998:761-769.
  • 3Trigwell S,Hayden RD and Nelson KF,et al.Effects of surface treatment on the surface chemistry of NiTi alloy for biomedical applications.Surface and Interface Analysis,1998,26 (7):483-489.
  • 4Daniel J,Chwirut H,Semih Oktay and Tara Ryan.Nitinol interventional cardiology devices-FDA's perspective.Proceedings of the Second International Conference on Shape Memory and Superelastic Technologies,USA,California,1997:377-379.
  • 5Wang XX,Mao ZY,Cao ZW and Qing RY,et al.The oxide film and its effect on shape memory effect and biocompatibility in a TiNi alloy.Proceedings of the International Symposium on Shape Memory materials,Beijing China,1994:621-625.
  • 6Christine Trepanier,Maryam Tabrizian,L' Hocine Yahia,Luc Bilodeau and Dominigue L Piron.Effect of Modification of oxide layer on NiTi stent corrosion resistance.J Biomed Mqter Res,1998,43:433-440.
  • 7Lam DJ and Ellis DE.Electronic structure of lanthanum perovskites with 3d transition elements.Phys Rev,1980,B22:5730.
  • 8Hua YJ,Liu X,Meng CG and Yang DZ.The throretical study of O2Adsorption on NiTi(100) and (110) Surface.Journal of Wuhan University of Technology Materials Science,2003,18(2):6-10.
  • 9Nelly A,Kachurovskaya,Mikheeva EP and Zhidomirov GM.Cluster molecular modeling of strong interaction for VOx/TiO2 supported catalyst.Journal of Molecular Catalysis A:Chemical,2002,178:191-198.
  • 10Nambu T,Ezaki H,Yukawa H and Morinaga M.Electronic structure and hydriding property of titanium compounds with CsCl-type structure.Journal of Alloys and Compounds,1999,293-295:213-216.