摘要 研究了以K、Na、Ca、Mg为主要基体组分时,ICP-MS测定多元素时的干扰情况,结果表明在基体<0.1%,基体对测定结果无明显干扰。ICP-MS外标法对天然地表水,地下水中土壤,水系沉积物,植物进行测定,在水样测定中,盐分<500mg/L,外标法与内标法及AAN,ICP-AES测定结果一致,土壤,水系沉积物,植物测定,当基体<0.1%外标法与内标法测定值一致。 The matrix interference of K,Na,Ca,Mg on ICP-MS were studied.When the body solution〈0.1%,the body has no apparent interference to the detections of trace elecments in the natural water on the earth s surface,the soil of groundwater,sediments of water system and the plants by ICP-MS with external calibration.Salinity〈500mg/L,and the measurements have no obvieous difference between external calibration and internal calibration in the water.
出处 《光谱实验室》 CAS CSCD 北大核心 2009年第5期1120-1125,共6页 Chinese Journal of Spectroscopy Laboratory
关键词 电感耦合等离子体-质谱 外标法 内标法 ICP-MS External Calibration Internal Calibration
  • 相关文献

参考文献7

  • 1A.R戴特,A.L格雷主编.电感耦合等离子体质谱分析的应用[M].李金影,姚继军译.北京;原子能出版社,1998.
  • 2Thomas R.Pradical Guide To ICP-MS[M].李金英等译.北京:原子能出版社,2007.2001.4-2003.2.
  • 3王小菇主编.电感耦合等离子体质谱应用实例[M].北京:化学工业出版社,2005.
  • 4Tanner S D,Barnov V I,Vollkopf V.Adynamic Reaction Cell for Inductively Coupled Plasma Mass Spectrometry[J].J.Anal.At.Spectom,2000,15,1261-1269.
  • 5U.S.EPA.Method 1638-Daermination of Trece Elements in Ambient Waters by On-Lane Chelation[S].EPA 821-R-95033.US,EPA,1995.4.
  • 6U.S.EPA.Inductive Coupled Plasma Mass Spectrometry[M].EPA Method 6020.US,EPA,1998.1.
  • 7李其英,何光涛.李海萍等.动态反应池等离子质谱法测定天然物料中痕量Al,As,Cr,Fe,Zn,Mn,Se[C].美国:Pekin Elmer公司ICP-AES与ICP-MS用户会论文集,2008.4.26-29.