高剂量率Ir-192遥控腔内后装治疗宫颈癌的剂量分布测试 被引量:1

A STUDY ON THE MEASUREMENT OF HIGH DOSE RATE IR-192 SEED SOURCE DOSE DISTRIBUTI ON IN REMOTELY CONTROLLED AFTERLOADING TREATING CARCINOMA OF UTERUS OF CERVIX
下载PDF
导出
摘要 目的:验证HY-HDR18型遥控后装治疗机所用施用器周围的剂量分布。方法:用TLD(热释光剂量计)[LiF(MG,Cu,P)]测量并验证由HY-HDR18型后装治疗计划系统设计的针对宫颈癌的妇科治疗计划及施用器周围的剂量分布。结果:TLD测量值与TPS计算值偏差4.2±2.5%。结论:理论值与实测值接近。 Objective: To verify the dose distribution around the gynecology applicators determined by the Treatment Planning System (TPS) of HY HDR 18. Methods: The dose distribution around the gynecology applicators was measured by LiF(Mg,Cu,P) chips (TLD). Results: The variations between TLD measurement and calculation of gynecology treatment plan are 4.2±2.5%. Conclusion: In the gynecology treatment plan formed by TPS(of HY HDR 18), the calculation data conform to that of experiment.HY HDR is very suitable for clinical use.
出处 《天津医科大学学报》 1998年第3期236-238,共3页 Journal of Tianjin Medical University
基金 天津医科大学总医院科研基金
关键词 高剂量率 剂量分布 铱192 宫颈癌 放射疗法 治疗 High Dose Rate Dose Distribution LiF(Mg,Cu,P) Carcinoma of Uterus Cervix