80kV64排多层螺旋CT低辐射肺动脉成像 被引量:44

Minimizing radiation dose of pulmonary angiography with 80 kV 64-row multidetector CT
原文传递
导出
摘要 目的研究80kV管电压64排多层螺旋CT低辐射肺动脉成像的可行性。方法64名志愿者随机分为2组。观察组35例采用80kV管电压CT结合右头臂静脉作为团注追踪(bolustracking)监测点进行肺动脉成像,对照组29例采用120kV管电压,延迟时间采用团注测试峰值时间+0.7s,进行肺动脉成像。测量容积CT剂量指数(CTDIvol)、剂量长度乘积(DLP),计算加权CT剂量指数(CTDIw)和有效剂量(E),测量肺动脉强化后CT值和背景噪声,计算信噪比(SNR)和对比噪声比(CNR)。5分法对两组图像质量进行目测评分。对DLP、E、SNR、CNR进行t检验;采用Mann—Whitney U检验比较两组图像目测评分结果。结果观察组DLP和E分别为(146.5±7.6)mGv·cm和(2.5±0.1)mSv,对照组DLP和E分别为(313.4±13.5)mGy·cm和(5.3±0.2)mSv,两者差异有统计学意义(P〈0.001),观察组的辐射剂量低于对照组。图像质量量化评价:观察组和对照组的SNR分别为32.6±3.6和31.35±2.0,差异无统计学意义(P=0.089)。观察组和对照组CNR分别为28.5±3.4和27.6±1.1,差异无统计学意义(P=0.18)。目测评分结果:观察组为5分8例,4分26例,3分1例;对照组为5分5例,4分24例,差异无统计学意义(P=0.76)。结论80kV64排多层螺旋CT结合右头臂静脉作为bolus tracking监测点可以减少辐射剂量,同时不降低图像质量,是肺动脉造影的首选检查方法。 Objective To assess the feasibility of minimizing radiation doses using 80 kV 64-row muhidetector computed tomography on pulmonary angiography. Methods 64 volunteers were devided into 2 groups to undergo MDCT pulmonary angiography (collimation, 64 × 0.625 mm; pitch, 1. 204 ). The observed group consisting of 35 patients were for pulmonary angiography with 80 kV voltage, 300 mAs, O. 75 s/roation. The control group consisting of 29 patients were for pulmonary angiography with the standard tube voltage (120 kV), 200 mAs, 0.5 s/roation and time delay using the peak time on bolus test added 0.7 s. Volume computed tomography dose index (CTDIvol), dose length product (DLP), pulmonary vessel enhancement and back noise were quantified. Signal-noise-ratio ( SNR ), contrast-to-noise-ratio ( CNR ), weighted computed tomography dose index (CTDIw) and effective dose (E) were calculated. Results of the two protocols were compared by using t test. Two radiologists used five-point scale to subjectively score arterial enhancement and depiction of small arterial detail. The scores were compared with Mann-Whitney U test. Results The 80 kV protocol had a significantly lower DLP and E than the 120 kV protocol [ (146.5 ± 7.6) mGy'cm vs (313.4 ± 13.5) mGy·cm, P〈0.001] and [(2.5±0.1) mGy vs (5.3±0.2) mGy, P〈0.001], respectively. The 80 kV protocol and the 120 kV protocol had identical SNR(32.6 ±3.6 vs 31.3 ± 2.3; P = 0.089) and CNR (28.5 ± 3.4 vs 27.6 ± 1.1 ; P = 0.18). No significant difference was found between the two protocols on scores for arterial enhancement and depiction of small arterial detail ( P = 0.76). Conclusions 80 kV 64 slices MDCT combined with right brachiocephalic vein as the monitoring site for bolus tracking could be the first choice of pulmonary angiography. It can reduce the radiation dose without sacrificing the image quality.
出处 《中华放射医学与防护杂志》 CAS CSCD 北大核心 2009年第5期535-539,共5页 Chinese Journal of Radiological Medicine and Protection
关键词 辐射剂量 X线计算机体层摄影术 肺动脉造影 Radiation dose X-ray computed tomography Pulmonary angiography
  • 相关文献

参考文献16

  • 1Prolngo JD,Gilkeson RC,Diaz M,et al.CT pulmonary angiongraphy:a comparative analysis of the utilization patterns in emergency department and hospitalized patients between 1998 and 2003.AIR,2004,183(4):1093-1096.
  • 2Diederich S.Radiation dose in helical CT for detection of pulmonary embolism.Eur Radiol,2003,13(7):1491-1493.
  • 3Tack D,De Maertelaer V,Petit W,et al.Multl-detector row CT pulmonary angiography:comparison of standard-dose and simulated low-dose techniques.Radiology,2005,236(1):318-325.
  • 4Coche E,Vvnckier S,Octave-Priqnot M.Pulmonary embolism:radiation dose with multi-detector row CT and digital angiography for diagnosis.Radiology,2006,240(3):690-697.
  • 5Mayo JR.Radiation dose issues in longitudinal studies involving computed tomngraphy.Proc Am Thorac,2008,5(9):934-939.
  • 6Kalra MK,Rizzo SM,Novelline RA.Reducing radiation dose in emergency computed tomography with automatic exposure control techniques.Emerg Radiol,2005,11(5):267-274.
  • 7Lee CH,Coo JM,Ye HJ,et al.Radiation dose modulation techniques in the multidetector CT era:from basics to practice.Radiographic,2008,28(5):1451-1459.
  • 8刘建新,孙红霞,唐光健,蒋学祥.多层螺旋CT低剂量对比剂肺动脉成像[J].中国医学影像技术,2006,22(7):1012-1014. 被引量:23
  • 9Ertl-Wagner BB,Hoffmann RT,Bruning R,et al.Multi-detector row CT angiography of the brain at various kilovoltage settings.Radiology,2004,231(2):528-535.
  • 10Shrimton PC,Hillier MC,Lewis MA,et al.National survey of doses from CT in the UK:2003.Br J Radiol,2006,79(948):968-980.

二级参考文献9

  • 1蔡欣,邓宇.多层螺旋CT在肺栓塞诊断中的应用[J].中国医学影像技术,2005,21(4):562-564. 被引量:22
  • 2张燕,金征宇,张竹花,赵文敏,王沄,朱海峰,梁继祥,陆菁菁,薛华丹.多层螺旋CT对周围肺动脉显示能力的研究[J].临床放射学杂志,2005,24(10):879-884. 被引量:28
  • 3Remy-Jardin M,Mastora I,Remy J.Pulmonary emboius imaging with multislice CT[J].Radiol Clin North Am,2003,41(31):507-519.
  • 4Stock KW,Remy-Jardin M.Pulmonary embolism:treatment with thrombus fragmentation and local fibrinolysis with recombinant human tissue plasminogen activator[J].Cardiovasc Intervent Radiol,1997,20(5):364-368.
  • 5Woodard PK,Sostmen HD,MacFall JR,et al.Detection of pulmonary embolism:comparison of contrast-enhanced spiral CT and time-of-flight MR technique[J].J Thorac Imaging,1995,10(1):59-72.
  • 6Remy-Jardin M,Remy J.Spiral CT angiography of the pulmonary circulation[J].Radiology,1999,212(3):615-636.
  • 7Schoepf UJ,Holzknecht N,Helmberger TK,et al.Subsegmental pulmonary emboli:improved detection with thirrcollimation multidetector row spiral CT[J].Radiology,2002,222(2):483-490.
  • 8Beigelman C,Chartrang-Lefebvre C,Howarth N,et al.Pitfalls in diagnosis of pulmonary embolism with helical CT angiography[J].AJR,1998,171(3):579-585.
  • 9李坤成,杜冬泽,宋云龙,刘玉清.肺动脉血栓栓塞的MRI诊断价值[J].中华放射学杂志,1999,33(5):299-302. 被引量:48

共引文献22

同被引文献280

引证文献44

二级引证文献329

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部