期刊文献+

反馈式人工肛门括约肌的经皮能量传输系统

Transcutaneous Energy Transmission System for Reactive Artificial Anal Sphincter
下载PDF
导出
摘要 针对反馈式人工肛门括约肌的供能问题,研究经皮能量传输系统原理,并研制出经皮能量传输系统,使得反馈式人工肛门括约肌能够模拟实现人体排便功能。通过实验分析接收端负载以及两线圈位置偏移对传输效率的影响,验证了系统的传输性能。实验结果表明,研制的经皮能量传输系统在发射和接收线圈轴向距离为15mm条件下,能稳定传输超过400mW功率,传输效率可达10%,保证人工肛门括约肌的正常工作。 The principle of transcutaneous energy transmission was studied, and a transcutaneous energy transmission system was designed for the energy supply of reactive artificial anal sphincter, which enables the reactive artificial anal sphincter to realize the defecation function of human body. The experiments analyzed the relationship between transmission efficiency and the load of the receiving end, and the two coils' position displacement as well. More than 400mW could be steadily received in the experiments with transmission efficiency of 10% at the axial separation of 15mm, which would guarantee the normal work of artificial anal sphincter.
出处 《北京生物医学工程》 2009年第4期404-407,共4页 Beijing Biomedical Engineering
基金 国家自然科学基金(30800235)、教育部博士点基金新教师基金(20070248094)资助
关键词 反馈式人工肛门括约肌 经皮能量传输 传输效率 reactive artificial anal sphincter transcutaneous energy transmission transmission efficiency
  • 相关文献

参考文献7

二级参考文献34

  • 1周丁华,黎介寿,颜国正,李宁,沈航.消化道内窥镜机器人新型微驱动机构的研究[J].中国生物医学工程学报,2005,24(2):227-230. 被引量:2
  • 2李炎冬,徐加鹤.肛门失禁的评估与治疗[J].结直肠肛门外科,2006,12(4):263-266. 被引量:5
  • 3马官营,颜国正,于莲芝,施建.人体消化道内微机电系统线圈耦合系数分析[J].北京生物医学工程,2006,25(5):486-489. 被引量:4
  • 4王坤东,颜国正.蠕动式微机器人结肠镜系统及模型[J].上海交通大学学报,2006,40(11):1955-1959. 被引量:5
  • 5Ong N T, Beasley S W. Comparison of clinical methods for the assessment of continence after repair of high anorectal anomalies [ J ], Pediatric Surgery International, 1990,5 (4) :233 - 237,
  • 6Hassink E A M,Rieu P N M A,Brugman A T M,Festen C. Quality of life after operatively corrected high anorectal malformation:a long follow-up study of patients aged 18 years and older[ J]. Journal of Pediatric Surgery, 1994,29 (6) :773 - 776.
  • 7Ninan G K, Puri P, Levatorplasty using a posterior sagittal approach in secondary faecal incontinence [ J ], Pediatric Surgery International, 1994,9(1 -2) :17 -20.
  • 8施诚仁,吴晔明,潘伟华,等.生物反馈式人工泵式肛门括约肌装置[P].中国专利:CN2448319Y.
  • 9Takeshi Inoue,Toshi Hiro Nishimura, Masao Saito, Musuo Nakaoka. A transcutaneous energy transmission system for an artificial organ by using a novel resonant converter[ J]. IEEE Transactions on Power Electronics, 1992,7 (2) :261 - 269.
  • 10Erickson R W, Maksimovic D. Fundamentals of power electronics [ M ]. Massachusetts, USA, Kluwer Academic Publication ,2001,