期刊文献+

基于VLSI无线供能通信系统耦合线圈实验 被引量:2

Coupling coil experiment in a VLSI wireless power harvesting telemetry system
下载PDF
导出
摘要 基于VLSI无线供能通信系统是微型生物植入电子装置的核心部件。本文开展了基于VLSI的无线供能通信系统中的耦合线圈实验。首先从耦合线圈尺寸、通信速率、耦合线圈同轴距离和2ASK解调方法等因素分析对通信质量的影响,其次,采用改进Miller(Modified Miller)编解码数据,进行二进制幅移键控调制类型,对三种线圈尺寸、四种通信速率进行了误码率测试实验。分析实验结果得出,在耦合线圈同轴距离、体外耦合线圈尺寸固定时,最大有效通信速率固定时,本系统中的有效体内耦合线圈半径为10mm。 VLSI wireless power harvesting telemetry system is the core of microelectronic implant. This paper states the coupling coil experiment in a VLSI wireless power harvesting telemetry system. First, the factors influencing telemetry quality are analyzed, which includes coil size, baud rate, internal distance and demodulation method. Then coupling coil experiments were carried out. Three sizes of coils, four kinds of baud rates were tested, and the modified Miller code and 2ASK modulation were adopted in the experiments. Experiment results show that the effective implant coil radius is 10 mm when the coupling coil size, internal distance and maximum baud rate are fixed.
出处 《仪器仪表学报》 EI CAS CSCD 北大核心 2009年第11期2428-2432,共5页 Chinese Journal of Scientific Instrument
基金 "863"计划资助项目(2007AA042300) 重庆市自然科学基金(CSTC2007BB5147)资助项目
关键词 改进Miller编解码 体内耦合线圈 品质因素 modified Miller code implant coil quality factor bit error rate
  • 相关文献

参考文献9

  • 1PARISE R J. Future all-electric transportation communication and recharging via wireless power beam[J]. Proceeding of SPIE, 2001,4214(7): 171 - 179.
  • 2赵德春,彭承琳.无线内窥镜中高效电磁感应连接[J].清华大学学报(自然科学版),2008,48(9):1519-1521. 被引量:5
  • 3马官营,颜国正,何秀.基于电磁感应的消化道内微系统的无线供能[J].上海交通大学学报,2008,42(5):798-802. 被引量:17
  • 4SAUER CH, MILUTIN S ESTANAC' EVIC'M, CAUWENBERGHS G, et al. Power harvesting and telemetry in CMOS for implanted devices[J]. IEEE Transactions On Circuits And Systems, 2005,52(12):2605- 2613.
  • 5HMIDA G. B, DHIEB M, GHARIANI H, et al. Transcutaneous power and high data rate transmission lbr biomedical implants[C]. Int. Conf. of DTIS, 2006:374-378.
  • 6MOLLAZADEH M, MURARI K, SCHWERDT H, CAUWENBERGHS G, et al. Wireless multichannel acquisition of neuropotentials[C]. Biomedical Circuits and Systems Conference, 2008. BioCAS 2008. IEEE. 20- 22 Nov. 2008.49-52.
  • 7郑小林,牟宗霞,侯文生,吴小鹰,阴正勤,Jiang Yingtao,Biswajit Das,彭承琳.基于微线圈阵列的多道神经电刺激信号透皮传输的初步实验研究[J].仪器仪表学报,2009,30(5):1110-1116. 被引量:7
  • 8邱祖江,郭亚炜,杨莲兴.一种改进Miller编解码的实现方法[J].微电子学,2000,30(3):176-178. 被引量:7
  • 9MURARI K, SAUER CH M, STANAC' CEVI' C MMILUTIN S C G, et al. Wireless multichannel integrated potentiostat for distributed neurotransmitter sensing[C]. Proceedings of the 2005 IEEE Engineering in Medicine and Biology 27th Annual Conference Shanghai, China, September 1-4, 2005:7329-7332.

二级参考文献28

  • 1周丁华,黎介寿,颜国正,李宁,沈航.消化道内窥镜机器人新型微驱动机构的研究[J].中国生物医学工程学报,2005,24(2):227-230. 被引量:2
  • 2马官营,颜国正,于莲芝,施建.人体消化道内微机电系统线圈耦合系数分析[J].北京生物医学工程,2006,25(5):486-489. 被引量:4
  • 3王坤东,颜国正.蠕动式微机器人结肠镜系统及模型[J].上海交通大学学报,2006,40(11):1955-1959. 被引量:5
  • 4赵海涛,吕晓迎,王余峰,王志功.透过皮肤的电磁耦合[J].东南大学学报(自然科学版),2007,37(3):368-373. 被引量:5
  • 5沈振元 聂志泉.通信系统原理[M].西安:西安电子科技大学出版社,1997..
  • 6Clements M Z, Erwin S H. High-efficiency coupling-insensitive transcutaneous power and data transmission via an inductive link [J]. IEEE Transactions on Biomedical Engineering, 1990, 37(7) : 716 - 722.
  • 7Vandevoorde G, Puers R. Wireless energy transfer for stand-alone systems: A comparison between low and high power applicability [J]. Sensors and Actuators A-Physical, 2001, 92: 305-311.
  • 8Wang G X, Liu W T, Sivaprakasam M, et al. Design and analysis of an adaptive transcutaneous power telemetry for biomedical implants [J]. IEEE Transactions on Circuits and Systems Ⅰ-Regular Papers, 2005, 52(10): 2109- 2117.
  • 9Sauer C, Stanacevic M, Cauwenberghs G, et al. Power harvesting and telemetry in CMOS for implanted devices [J]. IEEE Transactions on Circuits and Systems Ⅰ-Regular Papers, 2005, 52(12): 2605-2613.
  • 10Yamamoto H, Hiroto Kita. Enteroscopy [J]. Journal of Gastroenterology, 2005, 40:555 - 562.